Arca
Skip to content

Preludeease Hydrophilic Sheath Introducer

Introducer, Catheter. 510(k) K140543, from Merit Medical Systems, Inc. (South Jordan, UT). Product code DYB, class 2.

As of Nov 17, 2026, K140543 is an FDA 510(k) held by Merit Medical Systems, Inc., substantially equivalent on Oct 16, 2014 after 226 days, with 1 recall tied to it in FDA's recall database.

  • Substantially equivalent: Oct 16, 2014
  • Review time: 226 days
  • Recalls: 1
  • Cited as a predicate by: 2 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The Merit PreludeEASE Hydrophilic Sheath Introducer is intended to provide access and facilitate the percutaneous introduction of various devices into veins and/or arteries, including but not limited to the radial artery, while maintaining hemostasis for a variety of diagnostic and therapeutic procedures.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K140543, newest first; each opens FDA's recall record
StartedReasonStatus
Dec 1, 2016Z-0949-2017Recalling dilators included with the 6F PreludeEASE Hydrophilic Sheath Introducer because the dilator tips were manufactured with an undersized inner diameter and therefore will not allow passage of an 0.021 (0.53 mm)…Terminated

Other DYB Clearances

The newest 510(k)s for the same product code; their median review was 71 days; see every DYB device

Record

Type
Traditional
Received
Mar 4, 2014
Review panel
Cardiovascular
Regulation
21 CFR 870.1340

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.