Arca
Skip to content
Tools
Tools

Access Vitamin B12 Assay, Access Ferritin Assay, Access Folate Assay, Access HYPERsensitive hTSH Assay, Power Express Sample Processing System Generic Connection Module and Access Immunoassay System Reagents

Radioassay, Vitamin B12. 510(k) K140496, from Beckman Coulter, Inc. (Brea, CA). Product code CDD, class 2.

As of Nov 18, 2026, K140496 is an FDA 510(k) held by Beckman Coulter, Inc., substantially equivalent on Sep 16, 2014 after 201 days, with 4 recalls tied to it in FDA's recall database.

  • Review time: 201 days
  • Recalls: 4
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The basic Power Express is an automated sample handling system which processes sample tubes from the pre- centrifugation, pre-sorting step to presentation of centrifuged and decapped samples into Generic or Personality Racks for specific instruments. The Power Express can be configured with optional software and hardware to allow processing of sample tubes on Generic Connection Instruments.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K140496, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jul 17, 2024Z-2979-2024, Z-2980-2024, Z-2978-2024The adhesive joint between the plastic socket and metal component of the Power Express Sample Holder can fail, causing the metal part to detach and enter the centrifuge. This detachment can result in damage to the…Open
May 17, 2019Z-1741-2020Potential exposure to biohazard. Software design problem causes excess speed and vibrations in the unload arm movement which causes caused sample splashing.Terminated

Other CDD clearances

The newest 510(k)s for the same product code; see every CDD device

Record

Type
Special
Received
Feb 27, 2014
Review panel
Clinical Chemistry
Regulation
21 CFR 862.1810

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.