Arca
Skip to content
Tools
Tools

Integra External Fixation System

Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component. 510(k) K140463, from Ascension Orthopedic (Providence, UT). Product code KTT, class 2.

As of Nov 18, 2026, K140463 is an FDA 510(k) held by Ascension Orthopedic, substantially equivalent on Oct 10, 2014 after 228 days, with 3 recalls tied to it in FDA's recall database.

  • Review time: 228 days
  • Recalls: 3
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

INDICATIONS FOR USE The Integra External Fixation System is an external fixation device intended for use in the treatment of bone conditions including leg lengthening, osteotomies, arthrodesis, fracture fixation, and other bone conditions amenable to treatment by use of the external fixation modality.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K140463, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jul 20, 2016Z-2387-2016, Z-2386-2016Through the investigation of reported complaints Integra verified that there have been instances where the Universal Wire Fixation Bolts the Slotted Post broke during use at the surgical site.Terminated
Jul 1, 2016Z-2296-2016The silicone insert embedded in the rocker bottom has the potential to become dislodged or loose from the rocker bottom mold.Terminated

Other KTT clearances

The newest 510(k)s for the same product code; their median review was 84 days; see every KTT device

Record

Type
Traditional
Received
Feb 24, 2014
Review panel
Orthopedic
Regulation
21 CFR 888.3030

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.