Soundstar Eco 8F Gultrasound Catheter, Soundstar Eco 8F Ultrasound Catheter

Catheter, Ultrasound, Intravascular. 510(k) K140318, from Biosense Webster (Diamond Bar, CA). Product code OBJ, class 2.

As of Nov 20, 2026, K140318 is an FDA 510(k) held by Biosense Webster, substantially equivalent on Mar 12, 2014 after 30 days, with 1 recall tied to it in FDA's recall database.

  • Review time: 30 days
  • Recalls: 1
  • Cited as a predicate by: 3 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Biosense Webster SOUNDSTAR ® eco Diagnostic Ultrasound Catheter and related accessory devices are indicated for intra-cardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart. When used with compatible CARTO ® 3 EP Navigation Systems, the SOUNDSTAR ® eco Catheter provides location information.

Cited as a predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K140318, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
StartedRecallStatus
Sep 10, 2015Z-0003-2016Image disappeared from the cardiac ultrasound system when the CARTO 3 EP Navigation System needed restarting while the patient was experiencing pericardial effusion. Affects the CARTOSOUND Module of the CARTO 3 EP…Terminated

Other OBJ clearances

The newest 510(k)s for the same product code; their median review was 111 days; see every OBJ device

Record

Type
Special
Received
Feb 10, 2014
Review panel
Cardiovascular
Regulation
21 CFR 870.1200

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.