Arca
Skip to content
Tools
Tools

Tsp Hip Fracture Plating System

Device, Fixation, Proximal Femoral, Implant. 510(k) K140018, from Biomet, Inc. (Warsaw, IN). Product code JDO, class 2.

As of Nov 18, 2026, K140018 is an FDA 510(k) held by Biomet, Inc., substantially equivalent on Apr 4, 2014 after 91 days, with 2 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Apr 4, 2014
  • Review time: 91 days
  • Recalls: 2
  • Cited as a predicate by: 1 device

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The Biomet Hip Fracture Plating Systems are indicated for use in the open reduction and internal fixation of a wide variety of fractures of the proximal femur: intracapsular fractures and intertrochanteric fractures. Fracture plates with six or more femoral shaft fixation screw holes are also indicated for subtrochanteric fractures as well as proximal femoral osteotomies.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K140018, newest first; each opens FDA's recall record
Jan 7, 2020Z-1347-2020Potential increase in fracture rate was detected during internal testing, in addition to potential discoloration inside the inner pouch of the packaging.Terminated
Jun 10, 2016Z-2240-2016The screw may fracture at the junction of the shaft and the head during implantation. The patient may retain a foreign body if the fractured screw cannot be removed. Soft tissue irritation will not occur as the retained…Terminated

Other JDO Clearances

The newest 510(k)s for the same product code; their median review was 115 days; see every JDO device

Record

Type
Traditional
Received
Jan 3, 2014
Review panel
Orthopedic
Regulation
21 CFR 888.3030

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.