Flower Rear Foot Plating Set, Flower Calcaneus Planting Set, Flower Rearfoot Plating Set
Plate, Fixation, Bone. 510(k) K133930, from Flower Orthopedics Corporation (Horsham, PA). Product code HRS, class 2.
As of Nov 17, 2026, K133930 is an FDA 510(k) held by Flower Orthopedics Corporation, substantially equivalent on Mar 27, 2014 after 94 days, with 6 recalls tied to it in FDA's recall database.
- Substantially equivalent: Mar 27, 2014
- Review time: 94 days
- Recalls: 6
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe Flower Rearfoot Plating Set is intended to be used for internal fixation of fractures and reconstructions of bones of the rearfoot, including the calcaneus. Examples of these internal fixations and reconstructions include, but are not limited to extra-articular fractures, intra-articular fractures, joint depression fractures, tongue type fractures, severely comminuted fractures and…
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Synthes Locking Calcaneal PlatesK991407 · Synthes (Usa) · Jun 11, 1999
- Vlp Foot Talus Plates, Vlp Percutaneous Calcaneus Plates And Misc Vlp Bone Screws,…K110670 · Smith & Nephew, Inc. · Jul 12, 2011
- Smith & Nephew, Inc. Vlp Foot Plating, Screw System And AccessoriesK090675 · Smith & Nephew, Inc. · Jun 4, 2009
- Vlp 2.7MM Extra Large Percutaneous Calcaneus Plate, Peri-Loc 3.5MM Anterior Ankle Fusion…K120667 · Smith & Nephew, Inc. · Apr 4, 2012
- Flower Small And Medium Bone ImplantsK123562 · Flower Orthopedics Corporation · Mar 29, 2013
- Flower Small And Medium Implant SetK131657 · Flower Orthopedics Corporation · Aug 22, 2013
Recalls
Recalls FDA ties to K133930, newest first; each opens FDA's recall record| Jul 8, 2020 | Z-2783-2020The contract manufacturer of the calcaneus plate inadvertently used the incorrect material to manufacture the plates (a titanium alloy rather than pure titanium). | Terminated |
| Jul 8, 2020 | Z-2784-2020The contract manufacturer of the calcaneus plate inadvertently used the incorrect material to manufacture the plates (a titanium alloy rather than pure titanium). | Terminated |
| Jul 8, 2020 | Z-2785-2020The contract manufacturer of the calcaneus plate inadvertently used the incorrect material to manufacture the plates (a titanium alloy rather than pure titanium). | Terminated |
| Jul 8, 2020 | Z-2782-2020The contract manufacturer of the calcaneus plate inadvertently used the incorrect material to manufacture the plates (a titanium alloy rather than pure titanium). | Terminated |
| Jul 8, 2020 | Z-2787-2020The contract manufacturer of the calcaneus plate inadvertently used the incorrect material to manufacture the plates (a titanium alloy rather than pure titanium). | Terminated |
| Jul 8, 2020 | Z-2786-2020The contract manufacturer of the calcaneus plate inadvertently used the incorrect material to manufacture the plates (a titanium alloy rather than pure titanium). | Terminated |
Other HRS Clearances
The newest 510(k)s for the same product code; their median review was 79 days; see every HRS device- Arthrex Plates, Screws and Pins, and WashersK263602 · Arthrex, Inc. · Oct 29, 2026
- Trochlear Salvage Plating SystemK261135 · Skeletal Dynamics, Inc. · Oct 26, 2026
- ANTHEM Fracture SystemK262779 · Globus Medical, Inc. · Oct 23, 2026
- Tempo Plating SystemK263407 · Pace Surgical · Oct 23, 2026
- Park's Pectus SystemK263269 · Tdm Co., Ltd. · Oct 15, 2026
- TDM Large Bone Plate and Screw SystemK261611 · Tdm Co., Ltd. · Oct 2, 2026
- Alians ForearmK262649 · Newclip Technics · Sep 28, 2026
- ProRib Fixation SystemK261892 · Costa Surgical, Inc. · Sep 23, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Dec 23, 2013
- Review panel
- Orthopedic
- Regulation
- 21 CFR 888.3030
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files