Gas-Machine, Anesthesia. 510(k) K133886, from Draeger Medical GmbH (Andover, MA). Product code BSZ, class 2.

As of Nov 19, 2026, K133886 is an FDA 510(k) held by Draeger Medical GmbH, substantially equivalent on Jun 20, 2014 after 182 days, with 3 recalls tied to it in FDA's recall database.

  • Review time: 182 days
  • Recalls: 3
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

K133886 The anesthesia workstation Perseus A500 is intended for use in anesthetizing adults, children, and neonates and can be used for automatic and manual ventilation, pressure-supported spontaneous breathing, and spontaneous breathing. Perseus A500 is equipped with airway monitoring, gas measurement and device monitoring, O_{2 insufflation, and an anesthetic gas receiving system.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate
  • AtlanK230931 · Drägerwerk AG Co. Kgaa · Jul 23, 2023

Recalls

Recalls FDA ties to K133886, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Mar 20, 2024Z-1512-2024Unexpected shutdown while operating on battery power.Open
May 24, 2018Z-2254-2018The Draeger anesthesia device may be able to dose 100% N2O. In the event of a fault, the S-ORC module would not prevent setting an N2O flow that would result in a hypoxic mixture from being dosed to the patient.…Terminated
Nov 12, 2015Z-0360-2016There is a potential problem with the device main power switch of the Perseus A500 Anesthesia Workstation. The device power switch may spontaneously fail during use, causing one of the following to occur: 1. The…Terminated

Other BSZ clearances

The newest 510(k)s for the same product code; their median review was 265 days; see every BSZ device

Record

Type
Traditional
Received
Dec 20, 2013
Review panel
Anesthesiology
Regulation
21 CFR 868.5160

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.