The Kinegraph Vma (Vertebral Motion Analyzer) Software Version 2.2, The Motion Normalizer Patient Handling and Data

System, Image Processing, Radiological. 510(k) K133875, from Ortho Kinematics, Inc. (Washington, DC). Product code LLZ, class 2.

As of Nov 22, 2026, K133875 is an FDA 510(k) held by Ortho Kinematics, Inc., substantially equivalent on Jan 17, 2014 after 28 days, with 4 recalls tied to it in FDA's recall database.

  • Review time: 28 days
  • Recalls: 4
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The KineGraph VMATM software is a quantitative imaging software application intended to be used to process digital image files. It is designed for physicians and clinical professionals who are interested in the analysis of motion in medical images, particularly in musculoskeletal images of the spine.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K133875, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
May 23, 2018Z-2889-2018This email is to provide notification that, due to a software bug that has been corrected, the Radiological Read Report for the VMA test(s) contained an error.Open
Mar 2, 2017Z-2891-2018Ortho Kinematics notified customers that errors were contained in Radiological Read Report and VMA Report for the VMA test. The error resulted from an administrative error of VMA user account settings and has been fixed.Open
Jan 11, 2017Z-2882-2018Ortho Kinematics Inc. sent a Notice of Correction to Released Testing Results, Radiological Read Report for the Vertebral Motion Analyzer (VMA) test because it contained an error. The error occurred due to a software…Open
Aug 12, 2016Z-2890-2018Ortho Kinematics, Inc. notified customers of erroneous results of VMA tests due to a software bug that has been corrected.Open

Other LLZ clearances

The newest 510(k)s for the same product code; their median review was 145 days; see every LLZ device

Record

Type
Special
Received
Dec 20, 2013
Review panel
Radiology
Regulation
21 CFR 892.2050

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.