Ventilator, Continuous, Facility Use. 510(k) K133868, from Resmed, Ltd. (San Diego, CA). Product code CBK, class 2.

As of Nov 20, 2026, K133868 is an FDA 510(k) held by Resmed, Ltd., substantially equivalent on Jun 4, 2014 after 166 days, with 2 recalls tied to it in FDA's recall database.

  • Review time: 166 days
  • Recalls: 2
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Astral 100/150 provides continuous or intermittent ventilatory support for patients weighing more than 5kg (11 lb) who require mechanical ventilation. The Astral device is intended to be used in home, institution/hospital and portable applications for both invasive and non-invasive ventilation.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K133868, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jan 13, 2016Z-0746-2016Resmed Corporation is recalling Astral 100, 150 External Battery because of a start-up issue.Terminated
May 5, 2015Z-1750-2015The Astral device allows clinicians to disable all alarms including those that detect circuit disconnection. Inappropriate disabling of alarms in ventilator dependent patients can lead to hazards of insufficient…Terminated

Other CBK clearances

The newest 510(k)s for the same product code; their median review was 235 days; see every CBK device

Record

Type
Traditional
Received
Dec 20, 2013
Review panel
Anesthesiology
Regulation
21 CFR 868.5895

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files