Aquabplus,Aquabplus & Aquabplus B2,AQUABPLUS&AQUABPLUS Hf,Aquabplus& Aquabplus B2& Aquablus Hf

Subsystem, Water Purification. 510(k) K133829, from Vivonic GmbH (Sailauf, DE). Product code FIP, class 2.

As of Nov 18, 2026, K133829 is an FDA 510(k) held by Vivonic GmbH, substantially equivalent on Apr 22, 2014 after 126 days, with 3 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Apr 22, 2014
  • Review time: 126 days
  • Recalls: 3
  • Cited as a predicate by: 2 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The AquaBplus Water Purification Systems are reverse osmosis units intended for use with hemodialysis systems to remove organic and inorganic substances and microbial contaminants from the water used for treating hemodialysis patients or related therapies. These devices are intended to be a component in a complete water purification system, and are not complete water treatment systems.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

2 later 510(k) summaries name this device as a predicate
  • AquaBplus; AquaB LITEK232953 · Fresenius Medical Care Renal Therapies Group, LLC · Jun 13, 2024
  • AquaAK213507 · Fresenius Medical Care Rental Therapies Group, LLC · Jul 27, 2022

Recalls

Recalls FDA ties to K133829, newest first; each opens FDA's recall record
Aug 3, 2021Z-1364-2022AquaBplus HF module on the AquaBplus 2000 osmosis system, an error in the US Installation Guideline instructed that 3-ring mains, instead of 2-ring mains, may be installed in combination with an AquaBplus 2000 device.…Terminated
Mar 18, 2021Z-1536-2021AquaBplus HF module on the AquaBplus 2500 osmosis system, an error in the US Installation Guideline instructed that 3-ring mains, instead of 2-ring mains, may be installed in combination with an AquaBplus 2500 device.…Terminated
Sep 1, 2020Z-0098-2021Incorrectly crimped connectors were installed on the main protective earth conductor on these machines. If the machine is moved or maintenance is performed, these crimp connections could become loose and may adversely…Terminated

Other FIP Clearances

The newest 510(k)s for the same product code; their median review was 115 days; see every FIP device

Record

Type
Traditional
Received
Dec 17, 2013
Review panel
Gastroenterology, Urology
Regulation
21 CFR 876.5665

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.