Emprint Ablation System
System, Ablation, Microwave And Accessories. 510(k) K133821, from Covidien, LLC (Mansfield, MA). Product code NEY, class 2.
As of Nov 20, 2026, K133821 is an FDA 510(k) held by Covidien, LLC, substantially equivalent on Apr 28, 2014 after 133 days, with 7 recalls tied to it in FDA's recall database.
- Review time: 133 days
- Recalls: 7
- Cited as a predicate by: 9 devices
The Covidien Emprint™ Ablation System is intended for use in percutaneous, laparoscopic, and intraoperative coagulation (ablation) of soft tissue, including partial or complete ablation of non-resectable liver tumors. The Covidien Emprint™ Ablation System is not intended for use in cardiac procedures.
- Valleylab Microwave Ablation GeneratorK072687 · Covidien Lp, Formerly Know AS Valleylab, A Divison · Nov 25, 2008
- Certus 140 2.45GHZ Ablation SystemK113237 · Neuwave Medical, Inc. · Jan 4, 2012
- Emprint Ablation System with Thermosphere Technology (Overall System Name); Emprint Ablation Generator (CAGEN1);…K233838 · Medtronic, Inc. · Mar 28, 2024
- HEATfx Microwave Ablation System, HEATfx Microwave Ablation System Cart, HEATfx 2.1 Needle, HEATfx 2.1 S Needle,…K200061 · Galil Medical, Inc. · Jul 9, 2020
- Emprint Ablation System with Thermosphere Technology, Emprint SX Ablation Platform with Thermosphere TechnologyK193232 · Covidien, LLC · Feb 13, 2020
- Emprint Visualization ApplicationK192038 · Covidien, LLC · Dec 2, 2019
- Microwave Ablation SystemK183153 · Surgnova Healthcare Technologies (Zhejiang) Co., Ltd. · Jul 8, 2019
- Emprint Ablation Visualization ApplicationK180192 · Covidien, LLC · Mar 21, 2018
- Emprint SX Ablation Platform and AccessoriesK171358 · Coviden, LLC · Sep 1, 2017
- Emprint Ablation SystemK163105 · Covidien, LLC · Dec 6, 2016
- AveCure Microwave Coagulation-Ablation Generator, AveCure Microwave Large Antenna-Probe, AveCure Microwave Medium…K143203 · Medwaves Incorporated · Aug 26, 2015
| Aug 7, 2017 | Z-0560-2018, Z-0557-2018, Z-0561-2018, Z-0556-2018, Z-0558-2018, Z-0559-2018The firm received consumer reports of the ceramic trocar tip of the Emprint ablation antenna disengaging from the needle shaft post-ablation. This can result in the tip remaining in the patient. | Open |
| Feb 23, 2016 | Z-1187-2016Medtronic is voluntarily recalling specific item codes and production lots of Covidien Emprint" Percutaneous Antenna with Thermosphere" Technology due to the potential for the generator to shut down in the precence of… | Terminated |
- Microwave Ablation AntennasK261636 · Canyon Medical, Inc. · Oct 28, 2026
- Microwave Ablation DeviceK252632 · Nanjing Dewen Medical Technology Co., Ltd. · Mar 2, 2026
- Disposable Microwave Ablation Antenna (DW-XR-II1, DW-XR-II3, DW-XR-II4, DW-XR-II14, DW-XR-II5, DW-XR-II6, DW-XR-II7,…K253771 · Nanjing Devin Medical Technology Co., Ltd. · Mar 2, 2026
- swiftPro System (SWF-SPS); Swift System (SWF-SYS)K250718 · Emblation Limited · Aug 8, 2025
- Microwave Ablation Generator (KY-2000A, KY-2100A)K241827 · Canyon Medical, Inc. · Sep 26, 2024
- Microwave Ablation AntennasK241825 · Canyon Medical, Inc. · Sep 20, 2024
- IntelliBlate Microwave Ablation System Console (IB-CON); IntelliBlate Microwave Ablation System Mobile Cart (IB-CART);…K240480 · Varian Medical Systems, Inc. · Jul 19, 2024
- Temperature Sensor Probe (ABL-18TP20)K240784 · Varian Medical Systems, Inc. · May 29, 2024
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Dec 16, 2013
- Review panel
- General, Plastic Surgery
- Regulation
- 21 CFR 878.4400
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files