Somatom Force
System, X-Ray, Tomography, Computed. 510(k) K133589, from Siemens Medical Solutions USA, Inc. (Mavern, PA). Product code JAK, class 2.
As of Nov 17, 2026, K133589 is an FDA 510(k) held by Siemens Medical Solutions USA, Inc., substantially equivalent on Apr 17, 2014 after 146 days, with 9 recalls tied to it in FDA's recall database.
- Substantially equivalent: Apr 17, 2014
- Review time: 146 days
- Recalls: 9
- Cited as a predicate by: 3 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThis computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by the system can be used by a trained physician as an aid in diagnosis.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Somatom Definition FlashK121072 · Siemens Medical Systems, Inc. · May 8, 2012
Cited as a Predicate
3 later 510(k) summaries name this device as a predicate- Ziostation2K200315 · Ziosoft, Inc. · May 5, 2020
- SOMATOM Force, SOMATOM Definition Flash, SOMATOM Drive, SOMATOM Definition Edge, SOMATOM…K173630 · Siemens Medical Solutions USA, Inc. · Mar 30, 2018
- SOMATOM Force, SOMATOM Definition Flash, SOMATOM Definition Edge, SOMATOM Definition…K142955 · Siemens Medical Solutions USA, Inc. · Nov 24, 2015
Recalls
Recalls FDA ties to K133589, newest first; each opens FDA's recall record| Nov 24, 2020 | Z-0771-2021System does not trigger a cancel command followed by a reload of the scan with the new parameter(s). This issue may result in a delay in diagnosis and/or need for patient rescan | Terminated |
| Jun 22, 2020 | Z-3025-2020Sporadic problems with the current software may result in scanning workflow interruptions and unexpected user notifications causing delay in diagnosis or patient rescan | Terminated |
| Oct 7, 2019 | Z-0406-2020Software issue identified in the software versions syngo CT VB20 running on the SOMATOM CT Scanner: scan aborts and system crashes | Terminated |
| Jul 3, 2019 | Z-2274-2019Potential for four safety-related software issues impacting systems utilizing syngo.CT software version VB10A which may result in a scan abort. | Terminated |
| Oct 20, 2016 | Z-0692-2017Siemens is providing software update version VA50A_SP3 to address the software bugs that were identified through normal field monitoring and the Global Complaint Handling Process. Correction for the problems are as… | Terminated |
| Apr 1, 2016 | Z-1558-2016The Neonate Head protocol with the pediatric kernel Hp38 could result in artefacts and possibly lead to a misdiagnosis (i.e. either non-existing blood or liquid is mimicked in the images or actual existing blood or… | Terminated |
| Aug 18, 2015 | Z-0107-2016Software and firmware bugs | Terminated |
| Feb 6, 2015 | Z-1267-2015Possibility of image artifacts during data acquisition when using Adaptive Cardio Sequence, Turbo Flash, and Head modes. The update improves system start-up behavior and resuming, ECG handling with visual feedback of… | Terminated |
| Dec 3, 2014 | Z-0971-2015Artifacts are found in acquired imaging on the SOMATOM Force. | Terminated |
Other JAK Clearances
The newest 510(k)s for the same product code; their median review was 237 days; see every JAK device- AVIEW Aorta MetricsK260391 · Coreline Soft Co., Ltd. · Oct 16, 2026
- CAS-One IRK261225 · Cascination AG · Sep 29, 2026
- BodyTom 64 (NL4100); OmniTom Elite (NL5000)K261125 · Samsung HME America, Inc., dba NeuroLogica Corp. · Sep 22, 2026
- SCENARIA View Phase 6.0K261229 · Fujifilm Corporation · Sep 18, 2026
- FCT iStream Phase 2K253836 · Fujifilm Corporation · Sep 4, 2026
- BodyTom Elite (NL4000)K253450 · NeuroLogica Corp., a Samsung Company · Jul 28, 2026
- CardIQ SuiteK253770 · GE Medical Systems SCS · Jul 21, 2026
- Enhanced Boundary for PCCTK253560 · Ge Medical Systems, LLC · Jun 24, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Nov 22, 2013
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1750
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files