Appliance, Fixation, Spinal Interlaminal. 510(k) K133556, from Zimmer Spine, Inc. (Minneapolis, MN). Product code KWP, class 2.

As of Nov 17, 2026, K133556 is an FDA 510(k) held by Zimmer Spine, Inc., substantially equivalent on Jun 2, 2014 after 195 days, with 2 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Jun 2, 2014
  • Review time: 195 days
  • Recalls: 2
  • Cited as a predicate by: 6 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

When intended to promote fusion of the occipitocervical spine, cervical spine and the thoracic spine, (Occiput -T3), the Virage OCT Spinal Fixation System is indicated for the following: Degenerative disc disease (DDD) (neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, spinal stenosis, fracture, dislocation, failed…

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

6 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K133556, newest first; each opens FDA's recall record
Feb 3, 2016Z-1019-2016Zimmer Biomet Spine is initiating a Medical Device Correction regarding the Virage OCT Spinal Fixation System by providing additional guidance in the Virage OCT Surgical Technique Manual.Terminated
Nov 17, 2015Z-0625-2016Zimmer Biomet is initiating a recall of the Virage Occipital Eyelet, as it was found through product testing that the staking of the set screw was out of specification and can interfere with the insertion of the…Terminated

Record

Type
Traditional
Received
Nov 19, 2013
Review panel
Orthopedic
Regulation
21 CFR 888.3050

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files