Catheter, Percutaneous. 510(k) K133177, from Concentric Medical, Inc. (Mountain View, CA). Product code DQY, class 2.

As of Nov 20, 2026, K133177 is an FDA 510(k) held by Concentric Medical, Inc., substantially equivalent on Feb 25, 2014 after 131 days, with 3 recalls tied to it in FDA's recall database.

  • Review time: 131 days
  • Recalls: 3
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Concentric HD Guide Catheter is indicated for use in facilitating the insertion and guidance of an occlusion catheter, infusion catheter or other appropriate microcatheter into a selected blood vessel in the peripheral, coronary and neuro vascular systems. It may also be used as a diagnostic angiographic catheter and as a conduit for retrieval devices.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K133177, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Apr 23, 2024Z-1908-2024Neurovascular devices did not meet the minimum sample size required for endotoxin testing during manufacturing processes and are out of compliance with the AAMI ST72 standard, so these devices may exceed the acceptable…Open
Apr 3, 2024Z-1740-2024Stryker Neurovascular is recalling their DAC- Distal Access Catheter -038 136cm, a Distal Access Catheter by removal. The reason for the recall is: the DAC Lot #0000486382 was released with out-of-specification…Open
Aug 19, 2015Z-0116-2016One Distal Access Catheter intended for distribution in Korea, was found at a US location.Terminated

Other DQY clearances

The newest 510(k)s for the same product code; their median review was 42 days; see every DQY device

Record

Type
Traditional
Received
Oct 17, 2013
Review panel
Cardiovascular
Regulation
21 CFR 870.1250

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files