Draeger Jaundice Meter JM-105
Bilirubin (Total And Unbound) In The Neonate Test System. 510(k) K133175, from Draeger Medical Systems, Inc. (Telford, PA). Product code MQM, class 1.
As of Nov 27, 2026, K133175 is an FDA 510(k) held by Draeger Medical Systems, Inc., substantially equivalent on Nov 13, 2014 after 392 days, with 1 recall tied to it in FDA's recall database.
- Review time: 392 days
- Recalls: 1
The Jaundice Meter (JM-105) is a non-invasive transcutaneous bilirubinometer. It measures yellowness of subcutaneous tissue in newborn infants. The unit pro-vides a visual digital measurement that has been shown to correlate with serum bilirubin in newborn infants.
- Jaundice Meter, Model JM 103K042522 · Draeger Medical Infant Care, Inc. · Jul 6, 2005
| Started | Recall | Status |
|---|---|---|
| May 15, 2018 | Z-2047-2018Users have misinterpreted the display for out of range measurement indicated by the blinking "-0-" to mean a zero measurement | Terminated |
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Oct 17, 2013
- Review panel
- Clinical Chemistry
- Regulation
- 21 CFR 862.1113
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files