Arca
Skip to content
Tools
Tools

I2C

System, Radiation Therapy, Charged-Particle, Medical. 510(k) K132847, from Ion Beam Applications S.A. (Philadelphia, PA). Product code LHN, class 2.

As of Nov 20, 2026, K132847 is an FDA 510(k) held by Ion Beam Applications S.A., substantially equivalent on Nov 20, 2013 after 70 days, with 3 recalls tied to it in FDA's recall database.

  • Review time: 70 days
  • Recalls: 3

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

I2C is used with a charged particle or photon radiation therapy system for localization of the patient position with respect to the therapy equipment and to provide correction feedback to the radiation therapy device.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K132847, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Oct 28, 2016Z-2109-2018IBA is initiating this recall to address an issue identified with AdaPTinsight software and to reduce risk related to this problem.Terminated
Sep 16, 2016Z-2138-2018IBA is initiating this recall to notify its customers that the use of non-IBA accessories and/or third party equipment or unauthorized modifications to an IBA medical device may pose a safety risk.Terminated
Nov 7, 2014Z-0836-2015Two issues were identified: 1. In the release of adaPT insight v1.3.2., the kV/kV Single Source acquisition workflow was linked to a wrong geometrical calibration file, resulting in an alignment offset for this specific…Terminated

Other LHN clearances

The newest 510(k)s for the same product code; their median review was 190 days; see every LHN device

Record

Type
Traditional
Received
Sep 11, 2013
Review panel
Radiology
Regulation
21 CFR 892.5050

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files