Plate, Fixation, Bone. 510(k) K132594, from Solana Surgical, LLC (Memphis, TN). Product code HRS, class 2.

As of Nov 18, 2026, K132594 is an FDA 510(k) held by Solana Surgical, LLC, substantially equivalent on Dec 13, 2013 after 116 days, with 21 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Dec 13, 2013
  • Review time: 116 days
  • Recalls: 21
  • Cited as a predicate by: 1 device

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The Solana Surgical LLC, CrossCheck Plating System is indicated for stabilization and fixation of fresh fractures, revision procedures, joint fusion and reconstruction of small bones in the hand, feet, wrist, and ankles, fingers and toes. The system may be used in both pediatric and adult patients. The device is intended for single use.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K132594, newest first; each opens FDA's recall record
Jul 21, 2015Z-2542-2015These screws have a potential of passing through the plate in the case when torque is applied in excessive amounts, which may result in intraoperative malfunction.Terminated
Jul 21, 2015Z-2541-2015These screws have a potential of passing through the plate in the case when torque is applied in excessive amounts, which may result in intraoperative malfunction.Terminated
Mar 6, 2015Z-1494-2015The minor diameter, major diameter, thread pitch, and pitch on the Crosscheck are out of specification, which may result in intraoperative malfunctionTerminated
Mar 6, 2015Z-1483-2015The minor diameter, major diameter, thread pitch, and pitch on the Crosscheck are out of specification, which may result in intraoperative malfunction.Terminated
Mar 6, 2015Z-1490-2015The minor diameter, major diameter, thread pitch, and pitch on the Crosscheck are out of specification, which may result in intraoperative malfunctionTerminated
Mar 6, 2015Z-1498-2015The minor diameter, major diameter, thread pitch, and pitch on the Crosscheck are out of specification, which may result in intraoperative malfunctionTerminated
Mar 6, 2015Z-1500-2015The minor diameter, major diameter, thread pitch, and pitch on the Crosscheck are out of specification, which may result in intraoperative malfunction.Terminated
Mar 6, 2015Z-1487-2015The minor diameter, major diameter, thread pitch, and pitch on the Crosscheck are out of specification, which may result in intraoperative malfunctionTerminated
Mar 6, 2015Z-1497-2015The minor diameter, major diameter, thread pitch, and pitch on the Crosscheck are out of specification, which may result in intraoperative malfunctionTerminated
Mar 6, 2015Z-1489-2015The minor diameter, major diameter, thread pitch, and pitch on the Crosscheck are out of specification, which may result in intraoperative malfunctionTerminated
Mar 6, 2015Z-1492-2015The minor diameter, major diameter, thread pitch, and pitch on the Crosscheck are out of specification, which may result in intraoperative malfunctionTerminated
Mar 6, 2015Z-1484-2015The minor diameter, major diameter, thread pitch, and pitch on the Crosscheck are out of specification, which may result in intraoperative malfunctionTerminated
Mar 6, 2015Z-1486-2015The minor diameter, major diameter, thread pitch, and pitch on the Crosscheck are out of specification, which may result in intraoperative malfunctionTerminated
Mar 6, 2015Z-1496-2015The minor diameter, major diameter, thread pitch, and pitch on the Crosscheck are out of specification, which may result in intraoperative malfunctionTerminated
Mar 6, 2015Z-1482-2015The minor diameter, major diameter, thread pitch, and pitch on the Crosscheck are out of specification, which may result in intraoperative malfunction.Terminated
Mar 6, 2015Z-1499-2015The minor diameter, major diameter, thread pitch, and pitch on the Crosscheck are out of specification, which may result in intraoperative malfunction.Terminated
Mar 6, 2015Z-1495-2015The minor diameter, major diameter, thread pitch, and pitch on the Crosscheck are out of specification, which may result in intraoperative malfunctionTerminated
Mar 6, 2015Z-1493-2015The minor diameter, major diameter, thread pitch, and pitch on the Crosscheck are out of specification, which may result in intraoperative malfunctionTerminated
Mar 6, 2015Z-1485-2015The minor diameter, major diameter, thread pitch, and pitch on the Crosscheck are out of specification, which may result in intraoperative malfunctionTerminated
Mar 6, 2015Z-1488-2015The minor diameter, major diameter, thread pitch, and pitch on the Crosscheck are out of specification, which may result in intraoperative malfunctionTerminated
Mar 6, 2015Z-1491-2015The minor diameter, major diameter, thread pitch, and pitch on the Crosscheck are out of specification, which may result in intraoperative malfunctionTerminated

Other HRS Clearances

The newest 510(k)s for the same product code; their median review was 79 days; see every HRS device

Record

Type
Traditional
Received
Aug 19, 2013
Review panel
Orthopedic
Regulation
21 CFR 888.3030

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.