Discovery CT 590 RT, Optima CT 580, Optima CT 580 RT, Optima CT 580 W
System, X-Ray, Tomography, Computed. 510(k) K132410, from Ge Medical Systems, LLC (Waukesha, WI). Product code JAK, class 2.
As of Nov 20, 2026, K132410 is an FDA 510(k) held by Ge Medical Systems, LLC, substantially equivalent on Jan 17, 2014 after 168 days, with 2 recalls tied to it in FDA's recall database.
- Review time: 168 days
- Recalls: 2
- Discovery Model CT590 RT and Optima Model CT580 CT SystemsK093581 · Ge Medical Systems, LLC · Apr 19, 2011
- Advantage 4D OptionK032915 · Ge Medical Systems, LLC · Oct 30, 2003
| May 24, 2018 | Z-2581-2018Some CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply). | Terminated |
| May 19, 2017 | Z-2184-2017It was discovered that an issue has been identified that can result in a scan starting before the prescribed delay has completed. | Terminated |
- AVIEW Aorta MetricsK260391 · Coreline Soft Co., Ltd. · Oct 16, 2026
- CAS-One IRK261225 · Cascination AG · Sep 29, 2026
- BodyTom 64 (NL4100); OmniTom Elite (NL5000)K261125 · Samsung HME America, Inc., dba NeuroLogica Corp. · Sep 22, 2026
- SCENARIA View Phase 6.0K261229 · Fujifilm Corporation · Sep 18, 2026
- FCT iStream Phase 2K253836 · Fujifilm Corporation · Sep 4, 2026
- BodyTom Elite (NL4000)K253450 · NeuroLogica Corp., a Samsung Company · Jul 28, 2026
- CardIQ SuiteK253770 · GE Medical Systems SCS · Jul 21, 2026
- Enhanced Boundary for PCCTK253560 · Ge Medical Systems, LLC · Jun 24, 2026
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Aug 2, 2013
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1750
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files