Esoflip Balloon Dilation Catheter
Esophageal Dilator Balloon With Or Without Electrode Sensors. 510(k) K132337, from Crospon, Ltd. (Bonita Springs, FL). Product code PID, class 2.
As of Nov 24, 2026, K132337 is an FDA 510(k) held by Crospon, Ltd., substantially equivalent on Oct 7, 2013 after 73 days, with 1 recall tied to it in FDA's recall database.
- Review time: 73 days
- Recalls: 1
The EsoFLIP® Balloon Dilation Catheter is indicated for use in a clinical setting for dilating the gastro-esophageal junction of a patient with Achalasia.
- Wilson-Cook Balloon Dilation CathetersK900924 · Wilson-Cook Medical, Inc. · Dec 21, 1990
- Barostat Software OptionK120997 · Crospon, Ltd. · Aug 27, 2012
| Started | Recall | Status |
|---|---|---|
| Apr 25, 2024 | Z-2038-2024Due to saline conductivity issues, their is a potential that dilation catheters may provide inaccurate esophageal diameter measurements that may lead to various harms. | Open |
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Jul 26, 2013
- Review panel
- Gastroenterology, Urology
- Regulation
- 21 CFR 876.5980
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files