Arca
Skip to content

D-Rad Smart Pack

Plate, Fixation, Bone. 510(k) K132296, from Smith & Nephew, Inc. (Memphis, TN). Product code HRS, class 2.

As of Nov 17, 2026, K132296 is an FDA 510(k) held by Smith & Nephew, Inc., substantially equivalent on Jan 7, 2014 after 168 days, with 4 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Jan 7, 2014
  • Review time: 168 days
  • Recalls: 4
  • Cited as a predicate by: 4 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The D-RAD SMART PACK is intended for the fixation of fractures involving the distal radius.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

4 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K132296, newest first; each opens FDA's recall record
Mar 7, 2018Z-1317-2018One lot of D-RAD Self-Taping Locking Screws used with the Distal Radius Fracture Kit were machined out of specification. Screws measured over tolerance within the head thread form by 0.002-0.003 inch.Terminated
Mar 9, 2017Z-1857-2017The expiration dates on the labeling are incorrect, the actual expiration dates are before the date on the labeling.Terminated
Mar 9, 2017Z-1856-2017The expiration dates on the labeling are incorrect, the actual expiration dates are before the date on the labeling.Terminated
Mar 9, 2017Z-1855-2017The expiration dates on the labeling are incorrect, the actual expiration dates are before the date on the labeling.Terminated

Other HRS Clearances

The newest 510(k)s for the same product code; their median review was 79 days; see every HRS device

Record

Type
Traditional
Received
Jul 23, 2013
Review panel
Orthopedic
Regulation
21 CFR 888.3030

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.