D-Rad Smart Pack
Plate, Fixation, Bone. 510(k) K132296, from Smith & Nephew, Inc. (Memphis, TN). Product code HRS, class 2.
As of Nov 17, 2026, K132296 is an FDA 510(k) held by Smith & Nephew, Inc., substantially equivalent on Jan 7, 2014 after 168 days, with 4 recalls tied to it in FDA's recall database.
- Substantially equivalent: Jan 7, 2014
- Review time: 168 days
- Recalls: 4
- Cited as a predicate by: 4 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe D-RAD SMART PACK is intended for the fixation of fractures involving the distal radius.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Smith & Nephew Bone Plate System (Bone Plates, Bone Screws And Accessories)K993106 · Smith & Nephew, Inc. · Dec 9, 1999
- Peri-Loc Locking Bone Plates And Locking Bone Screws For The Upper ExtremityK051735 · Smith & Nephew, Inc. · Jul 19, 2005
Cited as a Predicate
4 later 510(k) summaries name this device as a predicate- Smith & Nephew, Inc. Plates and Screws Systems: EVOS, Peri-Loc, D-Rad, TC-100, VLP…K213126 · Smith & Nephew, Inc. · Sep 29, 2022
- Smith & Nephew Plates and ScrewsK210837 · Smith & Nephew, Inc. · Jun 29, 2021
- EVOS Wrist Fracture Plating SystemK181533 · Smith & Nephew, Inc. · Aug 9, 2018
- VLP Wrist Fracture SystemK161665 · Smith & Nephew, Inc. · Nov 15, 2016
Recalls
Recalls FDA ties to K132296, newest first; each opens FDA's recall record| Mar 7, 2018 | Z-1317-2018One lot of D-RAD Self-Taping Locking Screws used with the Distal Radius Fracture Kit were machined out of specification. Screws measured over tolerance within the head thread form by 0.002-0.003 inch. | Terminated |
| Mar 9, 2017 | Z-1857-2017The expiration dates on the labeling are incorrect, the actual expiration dates are before the date on the labeling. | Terminated |
| Mar 9, 2017 | Z-1856-2017The expiration dates on the labeling are incorrect, the actual expiration dates are before the date on the labeling. | Terminated |
| Mar 9, 2017 | Z-1855-2017The expiration dates on the labeling are incorrect, the actual expiration dates are before the date on the labeling. | Terminated |
Other HRS Clearances
The newest 510(k)s for the same product code; their median review was 79 days; see every HRS device- Arthrex Plates, Screws and Pins, and WashersK263602 · Arthrex, Inc. · Oct 29, 2026
- Trochlear Salvage Plating SystemK261135 · Skeletal Dynamics, Inc. · Oct 26, 2026
- ANTHEM Fracture SystemK262779 · Globus Medical, Inc. · Oct 23, 2026
- Tempo Plating SystemK263407 · Pace Surgical · Oct 23, 2026
- Park's Pectus SystemK263269 · Tdm Co., Ltd. · Oct 15, 2026
- TDM Large Bone Plate and Screw SystemK261611 · Tdm Co., Ltd. · Oct 2, 2026
- Alians ForearmK262649 · Newclip Technics · Sep 28, 2026
- ProRib Fixation SystemK261892 · Costa Surgical, Inc. · Sep 23, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Jul 23, 2013
- Review panel
- Orthopedic
- Regulation
- 21 CFR 888.3030
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files