Oec 9800 Plus

Interventional Fluoroscopic X-Ray System. 510(k) K132027, from Ge Oec Medical Systems, Inc. (Salt Lake City, UT). Product code OWB, class 2.

As of Nov 19, 2026, K132027 is an FDA 510(k) held by Ge Oec Medical Systems, Inc., substantially equivalent on May 1, 2014 after 304 days, with 8 recalls tied to it in FDA's recall database.

  • Review time: 304 days
  • Recalls: 8
  • Cited as a predicate by: 3 devices

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K132027, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jul 2, 2025Z-2341-2025GE HealthCare has become aware of a non-conformance to 21 CFR 1020.30 (k) in certain OEC 9800 and OEC 9900 Elite C-arm Systems. Specifically, impacted systems are missing the full layer of lead tape in the collimator…Open
Nov 21, 2024Z-0706-2025Due to manufacturing issue of insufficient sealing of X-ray tubes and can result in a potential oil leak and loss of fluoroscopic imaging capability.Open
May 15, 2023Z-2276-2023, Z-2279-2023Image intensifiers on systems, used to provide fluoroscopic/digital spot images can become detached if the systems encounter a large impact force when moved, which may result in the image intensifier falling, which…Open
May 8, 2020Z-2872-2020GEHC Surgery became aware that some users of the 9800 and 9900 fluoroscopic mobile C-Arms (referred to as the system here on) within the United States had requested that a Field Service Engineer (FSE) change the dose…Open
Dec 27, 2016Z-1059-2017GE Healthcare Surgery announces a voluntary field action for the OEC 9800 and OEC 9900 Elite due to additional supplementation to the user documentation to include areas for enhanced planned maintenance and component…Terminated
Dec 19, 2016Z-0973-2017GE Healthcare Surgery announces a voluntary field action for the OEC MiniView 6800, OEC 9900 Elite, and OEC 9800 because the workstation power cable assembly can result in intermittent power cord failures.Terminated
Oct 3, 2014Z-0195-2015GE Healthcare initiated a field correction because prior revision of components were potentially installed on certain OEC 9800, FlexiView 8800, OEC MiniView 6800 and OEC UroView 2800 systems.Terminated

Other OWB clearances

The newest 510(k)s for the same product code; their median review was 231 days; see every OWB device

Record

Type
Special
Received
Jul 1, 2013
Review panel
Radiology
Regulation
21 CFR 892.1650

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files