Airvo 2 Humidifier, Myairvo 2 Humidifier
Humidifier, Respiratory Gas, (Direct Patient Interface). 510(k) K131895, from Fisher & Paykel Healthcare, Ltd. (Manukau, NZ). Product code BTT, class 2.
As of Nov 18, 2026, K131895 is an FDA 510(k) held by Fisher & Paykel Healthcare, Ltd., substantially equivalent on Oct 24, 2013 after 121 days, with 2 recalls tied to it in FDA's recall database.
- Substantially equivalent: Oct 24, 2013
- Review time: 121 days
- Recalls: 2
- Cited as a predicate by: 9 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe AIRVO 2 and myAIRVO 2 humidifiers are for the treatment of spontaneously breathing patients who would benefit from receiving high flow warmed and humidified respiratory gases. This includes patients who have had upper airways bypassed. The flow may be from 2 – 60 L/min depending on the patient interface. The AIRVO 2 is for patients in hospitals and long-term care facilities.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Airvo Humidifier Myairo HumidifierK121975 · Fisher & Paykel Healthcare, Ltd. · Jan 3, 2013
Cited as a Predicate
9 later 510(k) summaries name this device as a predicate- F&P Optiflow Air/Oxygen Flow SourceK243917 · Fisher & Paykel Healthcare, Ltd. · Jul 14, 2025
- HFT150K233707 · Invent Medical Corporation · Jun 14, 2024
- BONHAWA Respiratory HumidifierK223863 · Telesair, Inc. · Oct 13, 2023
- F&P myAirvo 3K222292 · Fisher & Paykel Healthcare · May 8, 2024
- F&P Airvo 3K221338 · Fisher & Paykel Healthcare · Jan 27, 2023
- F&P 950 Respiratory HumidifierK220703 · Fisher & Paykel Healthcare · Jun 16, 2023
- TNI softFlow 50K180474 · Tni Medical AG · Oct 18, 2018
- SERVO-U Ventilator System version 2.1; SERVO-n Ventilator System version 2.1K180098 · Maquet Critical Care AB · Mar 19, 2019
- AirSpiral Heated Breathing TubeK162553 · Fisher & Paykel Healthcare · Dec 20, 2018
Recalls
Recalls FDA ties to K131895, newest first; each opens FDA's recall record| Sep 18, 2025 | Z-0158-2026When unintentionally disconnected from power source, humidifier device (used to deliver high flow warmed and humidified respiratory gases) may sound audible alert for less than 120 seconds, which may delay user… | Open, Classified |
| Apr 2, 2024 | Z-1823-2024Humidifier devices, used to deliver high flow respiratory therapy to patients, manufactured before 14 August 2017, have a speaker configuration that may result in distorted, intermittent or inaudible alarm sound levels.… | Open, Classified |
Other BTT Clearances
The newest 510(k)s for the same product code; their median review was 142 days; see every BTT device- Optiflow Oxygen Kit (AA484J)K262268 · Fisher and Paykel Healthcare Limited · Aug 27, 2026
- Hvt 2.0K254078 · Vapotherm, Inc. · Apr 13, 2026
- FL-10000U Respiratory HumidifierK251448 · Flexicare Medical Limited. · Jan 16, 2026
- Laboratorios Biogalenic Sterile Water for Inhalation, USPK251419 · Laboratorios Biogalenic S.A. DE C.V. · Jan 14, 2026
- Optiflow Switch+ Filtered Nasal Interface with CO2 Sampling (AA042J)K253479 · Fisher & Paykel Healthcare, Ltd. · Dec 19, 2025
- Hudson RCI® Comfort Flo Nasal Cannula Premature, Infant; Hudson RCI® Comfort Flo Plus and Soft Plus Cannula Extra SmallK250312 · Medline Industries, LP · Jul 2, 2025
- Optiflow+ Nasal Cannula - Small (OPT942); Optiflow+ Nasal Cannula - Medium (OPT944); Optiflow+ Nasal Cannula - Large…K251611 · Fisher & Paykel Healthcare, Ltd. · Jun 25, 2025
- Laboratorios Biogalenic Sterile Water for Inhalation, USPK242717 · Laboratorios Biogalenic S.A. DE C.V. · Jan 30, 2025
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Jun 25, 2013
- Review panel
- Anesthesiology
- Regulation
- 21 CFR 868.5450
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files