Evis Exera Iii Video System

Colonoscope And Accessories, Flexible/Rigid. 510(k) K131780, from Olympus Medical Systems Corp. (Center Valley, PA). Product code FDF, class 2.

As of Nov 21, 2026, K131780 is an FDA 510(k) held by Olympus Medical Systems Corp., substantially equivalent on Aug 29, 2013 after 73 days, with 8 recalls tied to it in FDA's recall database.

  • Review time: 73 days
  • Recalls: 8
  • Cited as a predicate by: 5 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

Endoscopes (GASTROINTESTINAL VIDEOSCOPE GIF-H190, GIF-HQ190, COLONOVIDEOSCOPE CF-HQ190L/I, CF-H190L/I, PCF-H190L/I, PCF-PH190L/I) This instrument is intended to be used with an Olympus video system center, endoscope position detecting unit (for CF-HQ190L/l), light source, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for…

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

5 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K131780, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jan 11, 2024Z-1210-2024Specific repaired colonovideoscopes were assembled without an adhesive which protects them from certain physical conditions such as high levels of vibration, fluctuation in temperature or severe shock loads.Open
Oct 16, 2023Z-0416-2024, Z-0418-2024, Z-0417-2024, Z-0419-2024, Z-0415-2024Specific devices may have an improperly repaired light guide connector due to missing adhesive to secure the suction channel mouthpiece to the scope connector. Issue may result in loss of image, decreased suction, or…Open
Sep 25, 2023Z-0341-2024Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for…Open
Sep 14, 2023Z-0056-2025Certain serial numbers of the CV-190 do not start up properly because parts that deviated from the specification were assembled into the power supply unit.Open

Other FDF clearances

The newest 510(k)s for the same product code; their median review was 178 days; see every FDF device

Record

Type
Special
Received
Jun 17, 2013
Review panel
Gastroenterology, Urology
Regulation
21 CFR 876.1500

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files