Au Bicarbonate Reagent
Enzymatic, Carbon-Dioxide. 510(k) K131546, from Beckman Coulter Ireland, Inc. (Brea, CA). Product code KHS, class 2.
As of Nov 25, 2026, K131546 is an FDA 510(k) held by Beckman Coulter Ireland, Inc., substantially equivalent on Oct 9, 2013 after 133 days, with 4 recalls tied to it in FDA's recall database.
- Review time: 133 days
- Recalls: 4
AU Bicarbonate reagent is intended for the quantitative determination of Bicarbonate in human serum and plasma on Beckman Coulter AU analyzers. Bicarbonate measurements are used in the diagnosis and treatment of numerous potentially serious disorders associated with changes in body acid-base balance acid-base balance. For In Vitro Diagnostic Use
- Carbon Dioxide - DstK960035 · Trace America, Inc. · Apr 8, 1996
| Oct 14, 2025 | Z-0997-2026Bicarbonate reagent may generate falsely high Bicarbonate results due to interference with Lactate Dehydrogenase (LDH) in the test sample. | Open |
| Apr 5, 2023 | Z-1606-2023, Z-1608-2023, Z-1607-2023There is a potential for a decrease in calibration ODs for Bicarbonate reagent which could result in calibration failures and/or quality control failures. | Open |
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- May 29, 2013
- Review panel
- Clinical Chemistry
- Regulation
- 21 CFR 862.1160
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files