Enzymatic, Carbon-Dioxide. 510(k) K131546, from Beckman Coulter Ireland, Inc. (Brea, CA). Product code KHS, class 2.

As of Nov 25, 2026, K131546 is an FDA 510(k) held by Beckman Coulter Ireland, Inc., substantially equivalent on Oct 9, 2013 after 133 days, with 4 recalls tied to it in FDA's recall database.

  • Review time: 133 days
  • Recalls: 4

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

AU Bicarbonate reagent is intended for the quantitative determination of Bicarbonate in human serum and plasma on Beckman Coulter AU analyzers. Bicarbonate measurements are used in the diagnosis and treatment of numerous potentially serious disorders associated with changes in body acid-base balance acid-base balance. For In Vitro Diagnostic Use

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K131546, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Oct 14, 2025Z-0997-2026Bicarbonate reagent may generate falsely high Bicarbonate results due to interference with Lactate Dehydrogenase (LDH) in the test sample.Open
Apr 5, 2023Z-1606-2023, Z-1608-2023, Z-1607-2023There is a potential for a decrease in calibration ODs for Bicarbonate reagent which could result in calibration failures and/or quality control failures.Open

Record

Type
Traditional
Received
May 29, 2013
Review panel
Clinical Chemistry
Regulation
21 CFR 862.1160

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc