Navigated Capstone Trials, Clydesdale Trials, Capstone & Clydesdale Inserter
Orthopedic Stereotaxic Instrument. 510(k) K131425, from Medtronic Navigation, Inc. (Memphis, TN). Product code OLO, class 2.
As of Nov 18, 2026, K131425 is an FDA 510(k) held by Medtronic Navigation, Inc., substantially equivalent on Aug 14, 2013 after 89 days, with 10 recalls tied to it in FDA's recall database.
- Substantially equivalent: Aug 14, 2013
- Review time: 89 days
- Recalls: 10
- Cited as a predicate by: 9 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe Navigated inserter is intended to be use during implant placement of Medtronic intervertebral body fusion devices during spinal surgery. The Navigated Trials are intended to be used to facilitate implant size selection of The navigated instruments are specifically designed for use with the StealthStation® System, which is indicated for any medical condition in which the use of stereotactic…
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Navigated Cd Horizon Solera Screwdrivers, Taps, Iliac Taps, Legacy TapsK124004 · Medtronic Navigation, Inc. · Mar 22, 2013
- Capstone Spinal SystemK073291 · Medtronic Sofamor Danek USA, Inc. · Apr 24, 2008
- Capstone R Spinal SystemK103731 · Medtronic Sofamor Danek · Jul 18, 2011
- Clydesdale Spinal SystemK083026 · Medtronic Sofamor Danek · Dec 29, 2008
- Clydesdales Spinal SystemK112405 · Medtronic Sofamor Danek USA, Inc. · Nov 21, 2011
Cited as a Predicate
9 later 510(k) summaries name this device as a predicate- Foundation Surgical Navigated Lateral Disc Prep InstrumentsK260011 · Foundation Surgical Group, Inc. · Apr 16, 2026
- Infinity™ OCT SystemK254165 · Medtronic Sofamor Danek USA, Inc. · Jan 21, 2026
- Stealth™ Spine Clamps; ModuLeX™ Shank MountsK242464 · Medtronic Navigation, Inc. · Jun 5, 2025
- Mazor X System (Mazor X Stealth Edition)K230064 · Mazor Robotics , Ltd. · Apr 7, 2023
- StealthStation Spinous Process ClampsK211442 · Medtronic Navigation · Jul 8, 2021
- Catalyft PL Expandable Interbody SystemK210425 · Medtronic Sofamor Danek USA, Inc. · May 24, 2021
- Navigated T2 STRATOSPHERE Inserters and Navigated TemplatesK191039 · Medtronic Sofamor Danek USA, Inc. · Aug 26, 2019
- Navigated ELEVATE Inserter, Navigated Disc Prep InstrumentsK163581 · Medtronic Sofamor Danek USA, Inc. · Apr 14, 2017
- Navigated Disc Prep Instruments and CAPSTONE TrialsK150231 · Medtronic Sofamor Danek USA, Inc. · Jun 16, 2015
Recalls
Recalls FDA ties to K131425, newest first; each opens FDA's recall record| Dec 4, 2024 | Z-1022-2025Due to an increase in complaint that the percutaneous reference pin would not fit into the patient reference frame or percutaneous pin adapter. | Open, Classified |
| Nov 28, 2023 | Z-0699-2024Their is the potential that the percutaneous pin may have a cross-pin that may render the pin unable to fit into tap cap, or too tight to remove. | Open, Classified |
| Nov 28, 2023 | Z-0698-2024Their is the potential that the percutaneous pin may have a cross-pin that may render the pin unable to fit into tap cap, or too tight to remove. | Open, Classified |
| Dec 6, 2018 | Z-0707-2020Under certain circumstances, the percutaneous pin adapter used with the StealthAir Frame Assembly may slightly rotate after surgical placement and not return to its original position, even when connections are tight and… | Terminated |
| Jun 29, 2017 | Z-3139-2017Potential for Spinal Clamps to be damaged when opened beyond intended limits, either no opening once attached to a patient's spinous process or leaving a component behind in the patient's body following breakage. | Terminated |
| Jun 29, 2017 | Z-3142-2017Potential for Spinal Clamps to be damaged when opened beyond intended limits, either no opening once attached to a patient's spinous process or leaving a component behind in the patient's body following breakage. | Terminated |
| Jun 29, 2017 | Z-3141-2017Potential for Spinal Clamps to be damaged when opened beyond intended limits, either no opening once attached to a patient's spinous process or leaving a component behind in the patient's body following breakage. | Terminated |
| Jun 29, 2017 | Z-3140-2017Potential for Spinal Clamps to be damaged when opened beyond intended limits, either no opening once attached to a patient's spinous process or leaving a component behind in the patient's body following breakage. | Terminated |
| Jun 29, 2017 | Z-3143-2017Potential for Spinal Clamps to be damaged when opened beyond intended limits, either no opening once attached to a patient's spinous process or leaving a component behind in the patient's body following breakage. | Terminated |
| Jul 18, 2016 | Z-2316-2016Medtronic Navigation, Inc. announces a voluntary field action for the Medtronic Navigation StealthStation Software applications affected by Neurologica BodyTom/CereTom floor-based scanners. | Terminated |
Other OLO Clearances
The newest 510(k)s for the same product code; their median review was 91 days; see every OLO device- GPS AnkleK262567 · Blue Ortho · Oct 26, 2026
- 7D Surgical System (10-0001)K260403 · 7D Surgical ULC · Oct 19, 2026
- TMINI Miniature Robotic SystemK263234 · THINK Surgical, Inc. · Oct 16, 2026
- CarboClear® Cervical Navigated InstrumentsK263154 · CarboFix Orhtopedics , Ltd. · Oct 9, 2026
- Spine Planning (3.0); Spine & Trauma Navigation; Alignment System SpineK262257 · Brainlab SE · Sep 30, 2026
- Remex Spine Surgery Navigation - Universal InstrumentsK262506 · Remex Medical Corp. · Aug 20, 2026
- Navbit MARQ (PRD-001-SUP (Supine Registration Kit), PRD-001-LAT (Lateral Registration Kit))K261410 · Suite 201, National Innovation Centre · Jul 30, 2026
- SPINEART Navigation Instrument SystemK261205 · Spineart SA · Jul 29, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- May 17, 2013
- Review panel
- Neurology
- Regulation
- 21 CFR 882.4560
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files