Arca
Skip to content
Tools
Tools

Navigated Capstone Trials, Clydesdale Trials, Capstone & Clydesdale Inserter

Orthopedic Stereotaxic Instrument. 510(k) K131425, from Medtronic Navigation, Inc. (Memphis, TN). Product code OLO, class 2.

As of Nov 18, 2026, K131425 is an FDA 510(k) held by Medtronic Navigation, Inc., substantially equivalent on Aug 14, 2013 after 89 days, with 10 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Aug 14, 2013
  • Review time: 89 days
  • Recalls: 10
  • Cited as a predicate by: 9 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The Navigated inserter is intended to be use during implant placement of Medtronic intervertebral body fusion devices during spinal surgery. The Navigated Trials are intended to be used to facilitate implant size selection of The navigated instruments are specifically designed for use with the StealthStation® System, which is indicated for any medical condition in which the use of stereotactic…

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

9 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K131425, newest first; each opens FDA's recall record
Dec 4, 2024Z-1022-2025Due to an increase in complaint that the percutaneous reference pin would not fit into the patient reference frame or percutaneous pin adapter.Open, Classified
Nov 28, 2023Z-0699-2024Their is the potential that the percutaneous pin may have a cross-pin that may render the pin unable to fit into tap cap, or too tight to remove.Open, Classified
Nov 28, 2023Z-0698-2024Their is the potential that the percutaneous pin may have a cross-pin that may render the pin unable to fit into tap cap, or too tight to remove.Open, Classified
Dec 6, 2018Z-0707-2020Under certain circumstances, the percutaneous pin adapter used with the StealthAir Frame Assembly may slightly rotate after surgical placement and not return to its original position, even when connections are tight and…Terminated
Jun 29, 2017Z-3139-2017Potential for Spinal Clamps to be damaged when opened beyond intended limits, either no opening once attached to a patient's spinous process or leaving a component behind in the patient's body following breakage.Terminated
Jun 29, 2017Z-3142-2017Potential for Spinal Clamps to be damaged when opened beyond intended limits, either no opening once attached to a patient's spinous process or leaving a component behind in the patient's body following breakage.Terminated
Jun 29, 2017Z-3141-2017Potential for Spinal Clamps to be damaged when opened beyond intended limits, either no opening once attached to a patient's spinous process or leaving a component behind in the patient's body following breakage.Terminated
Jun 29, 2017Z-3140-2017Potential for Spinal Clamps to be damaged when opened beyond intended limits, either no opening once attached to a patient's spinous process or leaving a component behind in the patient's body following breakage.Terminated
Jun 29, 2017Z-3143-2017Potential for Spinal Clamps to be damaged when opened beyond intended limits, either no opening once attached to a patient's spinous process or leaving a component behind in the patient's body following breakage.Terminated
Jul 18, 2016Z-2316-2016Medtronic Navigation, Inc. announces a voluntary field action for the Medtronic Navigation StealthStation Software applications affected by Neurologica BodyTom/CereTom floor-based scanners.Terminated

Other OLO Clearances

The newest 510(k)s for the same product code; their median review was 91 days; see every OLO device

Record

Type
Traditional
Received
May 17, 2013
Review panel
Neurology
Regulation
21 CFR 882.4560

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files