Arca
Skip to content

Calix Lumbar Spinal Implant System

Intervertebral Fusion Device With Bone Graft, Lumbar. 510(k) K131350, from X-Spine Systems, Inc. (Miamisburg, OH). Product code MAX, class 2.

As of Nov 17, 2026, K131350 is an FDA 510(k) held by X-Spine Systems, Inc., substantially equivalent on Oct 23, 2013 after 166 days, with 4 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Oct 23, 2013
  • Review time: 166 days
  • Recalls: 4
  • Cited as a predicate by: 3 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The X-spine Calix Lumbar Spinal Implant System is intended for spinal fusion procedures at one or two contiguous levels (L2 – S1 inclusive) in skeletally mature patients with degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) of the lumbosacral spine.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K131350, newest first; each opens FDA's recall record
May 17, 2016Z-0126-2017The trials and rasps used to prepare the surgical site for placement of the implant may become detached from the inserter assembly part and become lodged in the intervertebral space.Terminated
May 17, 2016Z-0127-2017The trials and rasps used to prepare the surgical site for placement of the implant may become detached from the inserter assembly part and become lodged in the intervertebral space.Terminated
May 17, 2016Z-0124-2017The trials and rasps used to prepare the surgical site for placement of the implant may become detached from the inserter assembly part and become lodged in the intervertebral space.Terminated
May 17, 2016Z-0125-2017The trials and rasps used to prepare the surgical site for placement of the implant may become detached from the inserter assembly part and become lodged in the intervertebral space.Terminated

Other MAX Clearances

The newest 510(k)s for the same product code; their median review was 142 days; see every MAX device

Record

Type
Traditional
Received
May 10, 2013
Review panel
Orthopedic
Regulation
21 CFR 888.3080

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.