Puritan Bennett 980 Series Ventilator System
Ventilator, Continuous, Facility Use. 510(k) K131252, from Covidien (Boulder, CO). Product code CBK, class 2.
As of Nov 17, 2026, K131252 is an FDA 510(k) held by Covidien, substantially equivalent on Feb 24, 2014 after 298 days, with 6 recalls tied to it in FDA's recall database.
- Substantially equivalent: Feb 24, 2014
- Review time: 298 days
- Recalls: 6
- Cited as a predicate by: 3 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe Puritan Bennett™ 980 Series Ventilator System is designed for use on patient population sizes from Neonatal (NICU) through Adult who require respiratory support or mechanical ventilation and weigh a minimum of 0.3kg (0.6lb).
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- 840 Ventilator System With Expanded Neomode OptionK092847 · Covidien, Formerly Nellcor Puritan Bennett, Inc. · Mar 11, 2010
- Avea VentilatorK103211 · Care Fusion · May 12, 2011
Cited as a Predicate
3 later 510(k) summaries name this device as a predicate- INOmax DSIR PlusK200389 · Mallinckrodt Manufacturing, LLC · Jun 17, 2020
- BEACON Caresystem Model 00002144K192584 · Mermaid Care A/S · Dec 18, 2019
- Puritan Bennett 980 Ventilator SystemK162738 · Covidien · Oct 27, 2017
Recalls
Recalls FDA ties to K131252, newest first; each opens FDA's recall record| Nov 21, 2017 | Z-1208-2018Rechargeable lithium-ion batteries with incorrect firmware that are used in certain Puritan Bennett 980 ventilators may not fully charge after installation. | Terminated |
| Jan 15, 2016 | Z-1181-2016Graphical user interface (GUI) unresponsive to touch and Loss of primary ventilation under certain circumstances. Covidien Respiratory and Monitoring Solutions, now a part of Medtronic, issued a field corrective action… | Terminated |
| Jul 16, 2015 | Z-2329-2015Reports in which tidal volumes reaching patients were lower than set tidal volumes in neonatal Volume Control Plus (VC+) Mode with active humidification. This situation may potentially lead to respiratory compromise if… | Terminated |
| Jan 12, 2015 | Z-1058-2015Covidien is issuing a voluntary field action for all Puritan Bennett 980 ventilators due to occasional GUI transient resets that last approximately 30 seconds. | Terminated |
| Oct 1, 2014 | Z-0180-2015Covidien is recalling certain Puritan Bennett 980 Ventilator Systems due to a blacklight inverter PCBA where a capacitor component can fail, causing a dim but functional screen and thermal odor. | Terminated |
| Oct 1, 2014 | Z-0112-2015A software issue may lead to ventilator inoperative situations. | Terminated |
Other CBK Clearances
The newest 510(k)s for the same product code; their median review was 235 days; see every CBK device- Evita (V800); Evita (V600)K253647 · Draegerwerk AG & CO Kgaa · May 7, 2026
- Inbentus VersatileK252663 · Iconic Solutions by Murcia, SL · Apr 14, 2026
- VPAP Pediatric Face MaskK251888 · Hsiner Co., Ltd. · Mar 12, 2026
- Vivo 45 LSK232365 · Oconnell Regulatory Consultants, Inc. · Dec 19, 2025
- F&P OptiNIV Hospital Vented Full Face Mask Compatible with Single-limb Circuits Size A…K252173 · Fisher & Paykel Healthcare, Ltd. · Sep 10, 2025
- SV600, SV800 VentilatorK243767 · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · Aug 28, 2025
- AVEA disposable expiratory filter/water trap (11790)K251133 · Telesair, Inc. · May 8, 2025
- AF531 Oro-Nasal SE Face MaskK243394 · Respironics, Inc. · Dec 30, 2024
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- May 2, 2013
- Review panel
- Anesthesiology
- Regulation
- 21 CFR 868.5895
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files