Amplatzer Torq Vue LP Delivery System
Catheter, Percutaneous. 510(k) K131063, from St Jude Medical (Plymouth, MN). Product code DQY, class 2.
As of Nov 17, 2026, K131063 is an FDA 510(k) held by St Jude Medical, substantially equivalent on Aug 15, 2013 after 121 days, with 1 recall tied to it in FDA's recall database.
- Substantially equivalent: Aug 15, 2013
- Review time: 121 days
- Recalls: 1
- Cited as a predicate by: 3 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe AMPLATZER ® TorqVue ® Low Profile Delivery System is indicated to provide a pathway through which devices are introduced within the chambers and coronary vasculature of the heart or in the peripheral vasculature.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Amplatzer Torqvue Delivery SystemK072313 · Aga Medical Corp. · Nov 2, 2007
- Amplatzer Torqvue Low Profile Delivery SystemK080757 · Aga Medical Corp. · Apr 18, 2008
Cited as a Predicate
3 later 510(k) summaries name this device as a predicate- Amplatzer TorqVue LP Delivery System; Amplatzer TorqVue LP CatheterK261825 · Abbott · Jul 2, 2026
- TriClip Steerable Guide CatheterK241538 · Abbott Medical · Aug 26, 2024
- TorqVue Low Profile Delivery CatheterK162228 · Aga Medical Corporation (Wholly Owned by St.Jude Medical Cor · Sep 9, 2016
Recalls
Recalls FDA ties to K131063, newest first; each opens FDA's recall record| Started | Reason | Status |
|---|---|---|
| Mar 17, 2025 | Z-1643-2025Device may have a small breach in the proximal end of the shaft under the strain relief of the delivery system. Breach is not visibly detectable and may lead to prolonged procedure, blood loss, or air ingress with… | Open, Classified |
Other DQY Clearances
The newest 510(k)s for the same product code; their median review was 42 days; see every DQY device- CrossFAST Integrated Microcatheter Advanced Delivery SystemK262817 · Vantis Vascular, Inc. · Sep 18, 2026
- NHanceITK260559 · IMDS Operation B.V. · Aug 14, 2026
- APRO 35 Swift CatheterK262392 · Alembic, LLC · Aug 12, 2026
- CPS Locator 3D Plus Delivery CatheterK261766 · CenterPoint Systems, LLC · Jul 29, 2026
- SSPC NXT Delivery CatheterK262240 · CenterPoint Systems, LLC · Jul 29, 2026
- Amplatzer TorqVue LP Delivery System; Amplatzer TorqVue LP CatheterK261825 · Abbott · Jul 2, 2026
- Amplatzer TorqVue Exchange SystemK261631 · Abbott Medical · Jun 17, 2026
- Perfect-O Ostial Positioning CatheterK260289 · Balosmark, Inc. · Jun 5, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Apr 16, 2013
- Review panel
- Cardiovascular
- Regulation
- 21 CFR 870.1250
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files