Cap, Device Disinfectant. 510(k) K130975, from Excelsior Medical Corp. (Washington, DC). Product code QBP, class 2.

As of Nov 23, 2026, K130975 is an FDA 510(k) held by Excelsior Medical Corp., substantially equivalent on Dec 24, 2013 after 260 days, with 2 recalls tied to it in FDA's recall database.

  • Review time: 260 days
  • Recalls: 2
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

SwabCap ® is intended for use on swab-able luer access valves as a cover to protect the luer access valves from potential contamination. The SwabCap ® acts as a physical barrier to contamination between line accesses and also serves as a disinfecting cleaner for use prior to line access.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF
  • SwabcapK083508 · Excelsior Medical Corp. · Apr 28, 2009

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K130975, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jun 16, 2025Z-2199-2025SwabFlush syringes are affected by the ICU Medical recall of their SwabCap Disinfecting Cap due to a manufacturing issue which affects the seal between the foil lid and plastic container of the Swab Cap. The issue may…Open
Jun 3, 2025Z-2116-2025Due to a manufacturing issue, disinfecting cap for needle-free connectors may have an incomplete seal between the foil lid and plastic container, which may result in isopropyl alcohol evaporation from the sponge, which…Open

Other QBP clearances

The newest 510(k)s for the same product code; their median review was 273 days; see every QBP device

Record

Type
Traditional
Received
Apr 8, 2013
Review panel
General Hospital
Regulation
21 CFR 880.5440

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files