Raystation
System, Planning, Radiation Therapy Treatment. 510(k) K130617, from Raysearch Laboratories AB (Stockholm, SE). Product code MUJ, class 2.
As of Nov 19, 2026, K130617 is an FDA 510(k) held by Raysearch Laboratories AB, substantially equivalent on May 10, 2013 after 64 days, with 6 recalls tied to it in FDA's recall database.
- Review time: 64 days
- Recalls: 6
- Cited as a predicate by: 2 devices
RayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user. The system functionality can be configured based on user needs.
- Proton VisionK000922 · Varian Assoc., Inc. · May 24, 2000
- .decimal Astroid Dosimetry AppK150547 · .Decimal, Inc. · May 15, 2015
- Ray StationK140187 · Raysearch Laboratories AB · May 15, 2014
| May 8, 2015 | Z-1679-2015RaySearch Laboratories AB became aware of an issue with the dose calculation for some imported VMAT plans where controls points are defined with wide gantry angle spacing. | Terminated |
| Apr 25, 2014 | Z-1806-2014This notice concerns a problem with the computation of ROI voxel volumes for cases where the CT has variable slice spacing. The dose grid volumetric representation of an ROI can be wrong. This affects all dose-volume… | Terminated |
| Feb 14, 2014 | Z-1481-2014RaySearch Laboratories has recalled "RaySearch Ray Station 2.0, 2.5, 3.0, 3.5 and 4.0" software due to dose miscalculations for bolus/external/fixation support structures on CT-image. | Terminated |
| Feb 6, 2014 | Z-1307-2014Potential for dose errors due to software program errors. | Terminated |
| Feb 6, 2014 | Z-1308-2014RaySearch Laboratories has recalled RaySearch Ray Station 4.0 software due to clinical dose calculation errors during radiation therapy. | Terminated |
| Oct 23, 2013 | Z-0281-2014Ray Search Laboratories found an issue with the dose calculation for Elekta motorized wedges in Ray Station 3.0, 3.5 and 4.0. The error presents when a wedge is used for a beam and the wedge angle differs from 60… | Terminated |
- RT Elements (4.6)K261655 · Brainlab SE · Oct 5, 2026
- Virtual Cone (1.0)K260750 · Varian Medical Systems, Inc. · Jul 8, 2026
- RadOncAIK253050 · Informai · Jun 18, 2026
- Multimodality Simulation Workspace (MM Sim) (v1.0.0)K260207 · Philips Medical Systems Nederland B.V. · May 27, 2026
- eXaSkin PlusK252784 · Anatomical Geometry S.L. · May 22, 2026
- ART.1-USK253078 · Seetreat Pty, Ltd. · May 14, 2026
- syngo.via RT Image Suite VC10K252304 · Varian Medical Systems, Inc. · Mar 18, 2026
- Monaco RTP System (6.3)K252002 · Elekta Solutions AB · Feb 19, 2026
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Mar 7, 2013
- Review panel
- Radiology
- Regulation
- 21 CFR 892.5050
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files