Lamp, Surgical. 510(k) K130513, from Maquet S.A.S. (Wayne, NJ). Product code FTD, class 2.

As of Nov 28, 2026, K130513 is an FDA 510(k) held by Maquet S.A.S., substantially equivalent on May 8, 2013 after 70 days, with 19 recalls tied to it in FDA's recall database.

  • Review time: 70 days
  • Recalls: 19

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

MAQUET TRIOP VOLISTA surgical lights are intended to be used to provide visible illumination of the surgical area or the patient during surgical operations, diagnostics and treatment.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K130513, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Nov 8, 2023Z-0758-2024A potential for a light system to fall in the operating room.Open
May 6, 2022Z-1308-2022Paint particles may detach and fall in the operating field. In the worst case scenario, infection, organ irritation, hypersensitivity reaction, and tissue granuloma may result if paint particles fall during surgical…Open
May 28, 2019Z-1862-2019, Z-1860-2019, Z-1853-2019, Z-1861-2019, Z-1857-2019, Z-1859-2019, Z-1856-2019, Z-1854-2019, Z-1864-2019, Z-1852-2019, Z-1858-2019, Z-1863-2019, Z-1855-2019Potential that the coating of the light head forks on the VOLISTA Surgical Lights may crack, posing the risk of particles falling into the surgical field during surgery that may result in patient infectionTerminated
Jun 14, 2018Z-2502-2018Certain VOLISTA StandOP Surgical Lights have been identified as having a potential light head detachment issue. Improper assembly of the bracket that connects the light head to the axle fork could result in the light…Terminated
Apr 19, 2018Z-1937-2018Potential keypad failure on VOLISTA StandOP Surgical Lights.Terminated
Feb 7, 2018Z-1349-2018The manufacturer received complaints indicating that the device's central handle holder detaches easily from the handle holder ring on the light head during clinical use. In the event of such detachment, particles from…Terminated
May 29, 2015Z-1990-2015Maquet Suspension Arms used with PowerLED/X-TEN/Volista Surgical Light Systems may grind at the junction of the suspension arms which can cause paint chips or particles to fall into the surgical field, when the…Terminated

Record

Type
Special
Received
Feb 27, 2013
Review panel
General, Plastic Surgery
Regulation
21 CFR 878.4580

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files