Orthopediatrics Acl Reconstructive System
Fastener, Fixation, Nondegradable, Soft Tissue. 510(k) K130217, from OrthoPediatrics Corp. (Warsaw, IN). Product code MBI, class 2.
As of Nov 17, 2026, K130217 is an FDA 510(k) held by OrthoPediatrics Corp., substantially equivalent on Jun 20, 2013 after 142 days, with 3 recalls tied to it in FDA's recall database.
- Substantially equivalent: Jun 20, 2013
- Review time: 142 days
- Recalls: 3
- Cited as a predicate by: 1 device
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe OrthoPediatrics ACL Reconstruction System is intended for fixation of tendons and ligaments during orthopedic reconstruction procedures such as Anterior Cruciate Ligament (ACL) Reconstruction.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Endobutton Continuous LoopK980155 · Smith & Nephew, Inc. · Apr 1, 1998
- Arthrex Interference ScrewK062466 · Arthrex, Inc. · Nov 29, 2006
Cited as a Predicate
1 later 510(k) summary names this device as a predicate- WasherCap Fixation SystemK212197 · Abanza Tecnomed S.L · Sep 22, 2022
Recalls
Recalls FDA ties to K130217, newest first; each opens FDA's recall record| Jun 5, 2020 | Z-2538-2020ShieldLoc implant component of the ACL Reconstruction Systemlots may be labeled with the incorrect size. Product labeled as 10-1008-4070 was found to have 10-1008-4060 product in the package. Product labeled as… | Terminated |
| Jul 25, 2014 | Z-2560-2014Weld fractures on the Sleeve Installation Tool for ACL reconstruction system. | Terminated |
| Nov 7, 2013 | Z-0579-2014Part # 01-1008-0040/0047, ACL Sleeve Installation/ Insertion Tool, shipped November 4, 2013 - November 6, 2013 may be etched with the wrong size. | Terminated |
Other MBI Clearances
The newest 510(k)s for the same product code; their median review was 30 days; see every MBI device- Arthrex FiberButtonK263477 · Arthrex, Inc. · Oct 27, 2026
- Q-FIX KNOTLESS All-Suture Anchor, 1.8mm, with one ULTRABRAID Suture, Blue; Q-FIX KNOTLESS…K263348 · Smith & Nephew, Inc. · Oct 23, 2026
- TeKBrace Knotless AnchorK263352 · TheraMicro · Oct 21, 2026
- Integrity Retractable Implant Inserter (6000127)K261869 · Anika Therapeutics, Inc. · Sep 17, 2026
- Aevumed SUPERCYNCH Button SystemK253065 · Aevumed, Inc. · Sep 2, 2026
- ArthroTAK Tendon Anchor KitK262809 · ArthroTAK, LLC · Sep 2, 2026
- Iconix Knotless TK262677 · Riverpoint Medical · Aug 27, 2026
- Grappler Suture Anchor SystemK262047 · Paragon 28, Inc. · Aug 3, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Jan 29, 2013
- Review panel
- Orthopedic
- Regulation
- 21 CFR 888.3040
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files