Disposable and Reusable Concentric Needle Electrodes,Disposable and Reusable Subdermal Needle Electrodes,
Electrode, Needle. 510(k) K130136, from Technomed Europe (Maastricht-Airport, NL). Product code GXZ, class 2.
As of Nov 20, 2026, K130136 is an FDA 510(k) held by Technomed Europe, substantially equivalent on Jul 19, 2013 after 182 days, with 2 recalls tied to it in FDA's recall database.
- Review time: 182 days
- Recalls: 2
- Cited as a predicate by: 3 devices
Needle Electrodes for Neurological purposes are intended for use with recording, monitoring equipment for the recording of biopotential signals including electroencephalograph (EEG), electromyograph (EMG) and nerve potential signals, and are intended for stimulation/recording with stimulation/recording equipment for electromyograph (EMG) and nerve potential signals.
- Electrode NeedlesK990015 · Technomed Europe · May 27, 1999
- Subdermal Needle ElectrodesK072276 · Xian Friendship Electronics Co., Ltd. · May 16, 2008
- Teca Elite Disposable Concentric NeedlesK112034 · Carefusion 209, Inc. · Aug 12, 2011
- Disposable Concentric Probe, Disposable Bipolar Probe, Disposable Monopolar ProbeK050325 · Technomed Europe · May 5, 2005
- Disposable Subdermal Needle Electrode, CorkscrewK241045 · Technomed Europe · Dec 11, 2024
- DMS Disposable Subdermal Needle ElectrodesK161566 · Daehan Medical Systems Co., Ltd. · Dec 21, 2016
- Disposable Concentric Needle electrodes, Disposable Monopolar Needle electrodes, Disposable EP Needle electrodes,…K152984 · Bio Protech, Inc. · Jun 3, 2016
| Aug 9, 2022 | Z-1737-2022Burns of varying degrees can occur to patients with electrodes left in while in the MRI scanner as a result of inadequate safety information in the IFU. | Terminated |
| Dec 8, 2021 | Z-0465-2022There is a high risk of the blue hub detaching from the corkscrew needle, resulting the the needle remaining in the patient's scalp. | Terminated |
- Disposable Subdermal Needle Electrode, CorkscrewK241045 · Technomed Europe · Dec 11, 2024
- Medical Disposable Sterile Needle ElectrodeK232581 · Suzhou Haishen Medical Device Associates Co., Ltd. · Dec 4, 2023
- Phantom XL Insulated DilatorsK231691 · TeDan Surgical Innovations, Inc. · Jun 29, 2023
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Jan 18, 2013
- Review panel
- Neurology
- Regulation
- 21 CFR 882.1350
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files