Revaclear 300 Dialyzer, Revaclear 400 Dialyzer

Dialyzer, High Permeability With Or Without Sealed Dialysate System. 510(k) K130039, from Gambro Renal Products, Inc. (Lakewood, CO). Product code KDI, class 2.

As of Nov 20, 2026, K130039 is an FDA 510(k) held by Gambro Renal Products, Inc., substantially equivalent on May 2, 2013 after 115 days, with 7 recalls tied to it in FDA's recall database.

  • Review time: 115 days
  • Recalls: 7

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The dialyzer is intended for the treatment of chronic and acute renal failure by hemodialysis.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K130039, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Nov 23, 2022Z-0764-2023, Z-0762-2023, Z-0763-2023Reports of Polyurethane Displacement in the dialyzer, and might be identified as a blood leakOpen
Nov 23, 2022Z-0760-2023Potential for Internal Blood Leaks due to a twisted gasket in the dialyzerOpen
Nov 15, 2019Z-1409-2020There is the potential presence of particular matter in the header caps of ten lots the Revaclear 300 Dialyzer.Terminated
May 31, 2019Z-1909-2019, Z-1908-2019There is a potential presence of ruptured dialyzer fibers which may lead to a blood leak during treatment.Terminated

Other KDI clearances

The newest 510(k)s for the same product code; their median review was 189 days; see every KDI device

Record

Type
Traditional
Received
Jan 7, 2013
Review panel
Gastroenterology, Urology
Regulation
21 CFR 876.5860

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files