Arca
Skip to content

Distal Ulna Locking Plate (Dul), Ulna Osteotomy Locking Plate (Uol), 3.0 Cancellous Locking Screw, 3.0 Cortical Locking

Plate, Fixation, Bone. 510(k) K130008, from I.T.S. GmbH (Prior Lake, MN). Product code HRS, class 2.

As of Nov 18, 2026, K130008 is an FDA 510(k) held by I.T.S. GmbH, substantially equivalent on Jun 25, 2013 after 174 days, with 1 recall tied to it in FDA's recall database.

  • Substantially equivalent: Jun 25, 2013
  • Review time: 174 days
  • Recalls: 1

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The Indications for Use of the I.T.S. GmbH - Distal Ulna Locking (DUL) and Ulna Osteotomy Locking (UOL) Plate Systems is to stabilize fractures/osteotomy in the long bone of the distal and mid-ulna of an adult patient. The Indications for Use of the I.T.S.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K130008, newest first; each opens FDA's recall record
StartedReasonStatus
Jan 23, 2026Z-1604-2026Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).Open, Classified

Other HRS Clearances

The newest 510(k)s for the same product code; their median review was 79 days; see every HRS device

Record

Type
Traditional
Received
Jan 2, 2013
Review panel
Orthopedic
Regulation
21 CFR 888.3030

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.