Arca
Skip to content

Acumed Cannulated Screw System

Screw, Fixation, Bone. 510(k) K123890, from Acumed, LLC (Hillsboro, OR). Product code HWC, class 2.

As of Nov 17, 2026, K123890 is an FDA 510(k) held by Acumed, LLC, substantially equivalent on Mar 18, 2013 after 90 days, with 2 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Mar 18, 2013
  • Review time: 90 days
  • Recalls: 2
  • Cited as a predicate by: 5 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The Acumed Cannulated Screw System consists of screws, washers, and accessories and is generally intended for fixation of fracture, fusion and osteotomies of large and small bones appropriate for size of device, which may include the following: Minimally invasive reconstruction of fractures and joints; Adjuvant for osteosynthesis in complex joint fractures; Multifragment joint fractures; Simple…

Cited as a Predicate

5 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K123890, newest first; each opens FDA's recall record
Mar 9, 2022Z-0958-2022Cannulated Screws 4.0mm x 10mm and 4.0mm x 12mm may not have sufficient axial pullout strength for the indicated use. Remote possibility of resulting in a mal or nonunion of the bone fragments, iatrogenic fracture,…Completed
Mar 9, 2022Z-0959-2022Cannulated Screws 4.0mm x 10mm and 4.0mm x 12mm may not have sufficient axial pullout strength for the indicated use. Remote possibility of resulting in a mal or nonunion of the bone fragments, iatrogenic fracture,…Completed

Other HWC Clearances

The newest 510(k)s for the same product code; their median review was 157 days; see every HWC device

Record

Type
Traditional
Received
Dec 18, 2012
Review panel
Orthopedic
Regulation
21 CFR 888.3040

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.