Arca
Skip to content
Tools
Tools

Biograph McT Family

System, Tomography, Computed, Emission. 510(k) K123737, from Siemens Medical Solutions USA, Inc. (Knoxville, TN). Product code KPS, class 2.

As of Nov 19, 2026, K123737 is an FDA 510(k) held by Siemens Medical Solutions USA, Inc., substantially equivalent on Jan 29, 2013 after 55 days, with 17 recalls tied to it in FDA's recall database.

  • Review time: 55 days
  • Recalls: 17
  • Cited as a predicate by: 3 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Siemens Biograph mCT systems are combined X-Ray Computed Tomography (CT) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K123737, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Apr 25, 2016Z-2000-2016, Z-1998-2016, Z-1999-2016, Z-1997-2016The bracket that is used to lift the cover has bolts that were not welded in potentially causing the cover to fall down during service.Terminated
Jul 14, 2015Z-2335-2015, Z-2331-2015, Z-2332-2015, Z-2334-2015, Z-2337-2015, Z-2336-2015, Z-2338-2015, Z-2333-2015Emission computed tomography systems and computed tomography xray systems with optional ceiling mounts may be missing a locking ring.Terminated
Nov 13, 2014Z-0811-2015, Z-0807-2015, Z-0809-2015, Z-0810-2015, Z-0808-2015Possibility for system display freeze during CT interventional procedures.Terminated

Other KPS clearances

The newest 510(k)s for the same product code; their median review was 88 days; see every KPS device

Record

Type
Special
Received
Dec 5, 2012
Review panel
Radiology
Regulation
21 CFR 892.1200

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files