Oec Brivo 865 Advance, Oec Brivo 785 Essential, and Oec Brivo 715 Prime

Image-Intensified Fluoroscopic X-Ray System, Mobile. 510(k) K123603, from Ge Healthcare Surgery GE Oec Medical Systems, Inc. (Salt Lake City, UT). Product code OXO, class 2.

As of Nov 20, 2026, K123603 is an FDA 510(k) held by Ge Healthcare Surgery GE Oec Medical Systems, Inc., substantially equivalent on May 17, 2013 after 177 days, with 5 recalls tied to it in FDA's recall database.

  • Review time: 177 days
  • Recalls: 5
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The OEC® Brivo® Mobile C-Arm X-Ray Products are designed to provide digital spot-film imaging and fluoroscopic image guidance for all adult and pediatric populations for orientations between patient anatomy and surgical instruments.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate
  • OEC OneK172700 · Ge Hualun Medical Systems Co. , Ltd. · Nov 9, 2017
  • OEC Elite MiniViewK160131 · Ge Hangwei Medical Systems Co., Ltd. · Jun 21, 2016

Recalls

Recalls FDA ties to K123603, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Sep 23, 2016Z-0071-2017GE Healthcare Surgery announces a voluntary field action for the OEC Brivo 715 Prime, OEC Brivo 685 Essential, and OEC Brivo 865 Advance because the circuit board that controls vertical motion of the C-arm may fail,…Terminated
Sep 15, 2014Z-0840-2015, Z-0841-2015mA Accuracy may exceed the design and labeling specification of+ 10%. Per 1020.32(f): Deviation of x-ray tube potential and current from the indicated values shall not exceed the maximum deviation as stated by the…Terminated
May 22, 2014Z-1638-2015Laser aimer assemblies were shipped without certification of the overall assembly or submitting a laser product report and therefore do not comply with the US regulations for laser products.Terminated
Sep 18, 2013Z-0066-2014GE is recalling certain lots of GE Brivo OEC715/785/865 C-Arm due to the potential for the wires in the high voltage cable bundle to fail, resulting in loss of system functionality.Terminated

Other OXO clearances

The newest 510(k)s for the same product code; their median review was 152 days; see every OXO device

Record

Type
Traditional
Received
Nov 21, 2012
Review panel
Radiology
Regulation
21 CFR 892.1650

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files