Arca
Skip to content
Tools
Tools

Olympus Energy Power System (Eps)

Electrosurgical, Cutting & Coagulation & Accessories. 510(k) K123429, from Gyrus Acmi, Inc. (Southborough, MA). Product code GEI, class 2.

As of Nov 19, 2026, K123429 is an FDA 510(k) held by Gyrus Acmi, Inc., substantially equivalent on Apr 5, 2013 after 149 days, with 12 recalls tied to it in FDA's recall database.

  • Review time: 149 days
  • Recalls: 12
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The system, including Image Guided Surgery compatibility, and optional Suction console, is intended for cutting, coagulation, drilling, debriding, and removal of bone, and soft and hard tissue in general ENT, Sinus / Rhinology, Nasopharyngeal / Laryngology, and Head & Neck procedures.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K123429, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Mar 8, 2019Z-1179-2019, Z-1186-2019, Z-1184-2019, Z-1181-2019, Z-1187-2019, Z-1182-2019, Z-1180-2019, Z-1178-2019, Z-1183-2019, Z-1185-2019Olympus Diego Elite Consoles may inadvertently permit activation of the RF energy feature when used with Diego Elite bladesTerminated
May 8, 2015Z-1752-2015, Z-1753-2015Potential lack of sterility assurance.Terminated

Other GEI clearances

The newest 510(k)s for the same product code; their median review was 167 days; see every GEI device

Record

Type
Traditional
Received
Nov 7, 2012
Review panel
General, Plastic Surgery
Regulation
21 CFR 878.4400

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files