Arca
Skip to content
Tools
Tools

Syngo.Via

System, Image Processing, Radiological. 510(k) K123375, from Siemens AG (Erlangen, DE). Product code LLZ, class 2.

As of Nov 19, 2026, K123375 is an FDA 510(k) held by Siemens AG, substantially equivalent on Nov 20, 2012 after 19 days, with 8 recalls tied to it in FDA's recall database.

  • Review time: 19 days
  • Recalls: 8
  • Cited as a predicate by: 4 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

syngo.via is a software solution intended to be used for viewing, manipulation, communication, and storage of medical images. It can be used as a stand-alone device or together with a variety of cleared and unmodified syngo based software options.

Cited as a predicate

4 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K123375, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Mar 27, 2018Z-1778-2018Functionality in the report sections "Findings Information" and "Summary of Measured Findings" not functioning properly. The corrected values modified in the report are neither saved, nor printed, nor sent to the…Terminated
Jan 10, 2018Z-1261-2018When the archiving configuration is changed, data received/created after the upgrade may be flagged as "Not to be archived". This is caused by the automatic function for cleaning up temporary data being disabled by the…Terminated
Jan 8, 2018Z-1303-2018A software functionality in the report sections Findings Information and Summary of Measured Findings is not performing properly. The system will use the original values instead of displaying and saving…Terminated
Aug 31, 2017Z-0808-2018A software update was released on August 22, 2017. Performing this software installation and upgrading syngo.via system to the software version VB20A_HF02 will resolve various bugs and inconsistent diagnosis with the…Terminated
Mar 6, 2017Z-1689-2017Software changes now available to address several issuesTerminated
Aug 5, 2016Z-2853-2016Software error. Incorrect values for the volume calculation from a freehand VOI at the customer site. In volume calculations of prostate as well as in volume calculations of liver were too high.Terminated
Jun 23, 2016Z-2245-2016Incorrect values for the volume calculation. Software update VB30B via Update Instructions SY018/16/P to resolve software errors.Terminated
May 15, 2015Z-1705-2015measurements drawn on the 2nd and subsequent images of the series are not visible on printouts when the series is sent to print.Terminated

Other LLZ clearances

The newest 510(k)s for the same product code; their median review was 145 days; see every LLZ device

Record

Type
Traditional, third-party review
Received
Nov 1, 2012
Review panel
Radiology
Regulation
21 CFR 892.2050

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files