Arca
Skip to content
Tools
Tools

Alignrt Plus

Accelerator, Linear, Medical. 510(k) K123371, from Vision RT Limited (London, GB). Product code IYE, class 2.

As of Nov 20, 2026, K123371 is an FDA 510(k) held by Vision RT Limited, substantially equivalent on Jan 8, 2013 after 68 days, with 5 recalls tied to it in FDA's recall database.

  • Review time: 68 days
  • Recalls: 5
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The AlignRT Plus system is indicated for use to position and monitor patients relative to the prescribed treatment isocentre, and to withhold the beam automatically during radiation delivery. For cranial treatments, a manual head adjuster is included which can be used in concert with AlignRT Plus to provide fine corrections for pitch, roll and yaw rotations.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K123371, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Aug 24, 2018Z-0065-2019RTInterface error in the event of an I/O error between the IGC PCB and the Framegrabber PCBTerminated
Jun 30, 2016Z-2667-2016Vision RT Ltd received three (3) reports from different sites that discovered the following issue, unintended changes were made to planned couch parameters, specifically the couch rotation parameter, during patient…Terminated
Jun 23, 2016Z-2150-2016, Z-2149-2016Varian Medical Systems has received reports that unintended changes were made to planned couch parameters, specifically the couch rotation parameter, during patient set-up. This can occur when a user sets a different…Terminated
Oct 15, 2014Z-0464-2015Potential failure of AlignRT to assert interlock.Terminated

Other IYE clearances

The newest 510(k)s for the same product code; their median review was 173 days; see every IYE device

Record

Type
Traditional
Received
Nov 1, 2012
Review panel
Radiology
Regulation
21 CFR 892.5050

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.