Arca
Skip to content
Tools
Tools

Nordic Tumorex

System, Image Processing, Radiological. 510(k) K123306, from Nordicneurolab (Bergen, Hordaland, NO). Product code LLZ, class 2.

As of Nov 18, 2026, K123306 is an FDA 510(k) held by Nordicneurolab, substantially equivalent on May 20, 2013 after 208 days, with 1 recall tied to it in FDA's recall database.

  • Review time: 208 days
  • Recalls: 1

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

nordicTumorEx is an image processing software package to be used by trained professionals including but not limited to physicians and medical technicians. The software runs on a standard "off-the-shelf" PC workstation and can be used to perform image viewing, processing and analysis of medical images. Data and images are acquired through DICOM compliant imaging devices and modalities.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K123306, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
StartedRecallStatus
Dec 15, 2014Z-2042-2018An error was discovered in the interpretation of certain DICOM header tags that may lead to incorrect orientation labeling, and thus and indirect left-right, up-down or anterior-posterior flipping of images.Terminated

Other LLZ clearances

The newest 510(k)s for the same product code; their median review was 145 days; see every LLZ device

Record

Type
Traditional
Received
Oct 24, 2012
Review panel
Radiology
Regulation
21 CFR 892.2050

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.