Centricity Pacs-Iw With Universal Viewer
System, Image Processing, Radiological. 510(k) K123174, from GE Healthcare (Barrington, IL). Product code LLZ, class 2.
As of Nov 20, 2026, K123174 is an FDA 510(k) held by GE Healthcare, substantially equivalent on Nov 16, 2012 after 37 days, with 10 recalls tied to it in FDA's recall database.
- Review time: 37 days
- Recalls: 10
- Cited as a predicate by: 6 devices
Centricity PACS-1W with Universal Viewer is a software only product that is part of a Picture Archiving and Communication System (PACS) medical device. Centricity PACS-IW with Universal Viewer is a device that displays medical images (including mammograms) and data from various imaging sources.
- Centricity Pacs-IwK121387 · Ge Healthcare Integrated IT Solutions · Jun 5, 2012
- RadUnity (Version 1.0.0)K242810 · Radunity Corp. · Nov 26, 2024
- Synaptive ImageDrive ProK153284 · Synaptive Medical, Inc. · Mar 29, 2016
- EV Insite SystemK152716 · Psp Corporation · Mar 24, 2016
- Centricity Universal ViewerK150420 · GE Healthcare · May 26, 2015
- VM Pacs With VM Medical WorkstationK141561 · Visual Medica · Jan 16, 2015
- Centricity Universal Viewer Zero Footprint ClientK131977 · GE Healthcare · Sep 17, 2013
| Jan 30, 2026 | Z-1571-2026There is a potential cybersecurity vulnerability affecting certain versions of Centricity Universal Viewer. User login credentials may be exposed on the local client workstation, which could allow an unauthorized… | Open |
| Feb 19, 2025 | Z-1484-2025There is a potential security vulnerability in Centricity Universal Viewer (UV), Centricity PACS-IW (PACS-IW), Centricity Radiology RA600 (RA600), & Centricity Cardiology CA1000 (CA1000) where the service login… | Open |
| Sep 14, 2023 | Z-0139-2024GE HealthCare has become aware of an issue where images from two different patients may be contained in a single study, when stored in Centricity PACS-IW V3.7.x, Centricity PACS-IW with Universal Viewer V5.0 and… | Open |
| Mar 28, 2023 | Z-1500-2023GE HealthCare has become aware of an issue where information from two different patients can be mismatched when correcting patient or study information. | Open |
| Oct 28, 2022 | Z-0721-2023There is the potential that after installation of the IAS tool, some studies can remain hidden without user notification. Access to those hidden studies requires GE Healthcare assistance, which can result in a delay in… | Open |
| Aug 6, 2021 | Z-2502-2021Image acquisition failures and synchronization failure with the Centricity Enterprise Archive | Open |
| May 17, 2019 | Z-1885-2019There is a potential that one or more images or image series may be missing from exams without a warning displayed in the viewer. | Terminated |
| Oct 11, 2018 | Z-0374-2019Synchronization failure | Terminated |
| Jan 25, 2018 | Z-2101-2018Potential that one or more image series may be missing from an exam without a user warning displayed in the Viewer. | Terminated |
| Sep 28, 2015 | Z-0225-2016Inaccurate distance measurements with magnified projection X-ray images. | Terminated |
- 3D Anatomical ModelK262132 · The Jacobs Institute · Nov 5, 2026
- LiLi DICOM Viewer (LiLi )K260454 · Elevate Imaging, Inc. · Nov 4, 2026
- Vitrea Advanced Visualization, Version 7.16K263648 · Canon Medical Informatics, Inc. · Oct 29, 2026
- MIM – KineticIDK261763 · Mim Software, Inc. · Oct 26, 2026
- PV-iCurve ProK261531 · Perception Vision Medical Technology Co., Ltd. · Oct 21, 2026
- Eden PACSK261788 · Higia., Inc. d/b/a Eden · Oct 21, 2026
- Syngo Carbon Space (VA4)K263048 · Siemens Healthineers AG · Sep 29, 2026
- Syneris PACS v.1.0.1.0K261874 · Syneris · Sep 23, 2026
- Document
- 510(k) summary (PDF) ↗
- Type
- Special
- Received
- Oct 10, 2012
- Review panel
- Radiology
- Regulation
- 21 CFR 892.2050
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files