Arca
Skip to content

Olympic Brainz Monitor

Amplitude-Integrated Electroencephalograph. 510(k) K123079, from Excel-Tech Ltd. (Xltek) (Oakville, CA). Product code OMA, class 2.

As of Nov 17, 2026, K123079 is an FDA 510(k) held by Excel-Tech Ltd. (Xltek), substantially equivalent on May 8, 2013 after 219 days, with 2 recalls tied to it in FDA's recall database.

  • Substantially equivalent: May 8, 2013
  • Review time: 219 days
  • Recalls: 2

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The Olympic Brainz Monitor (OBM) is a three channel electroencephalograph (EEG) acquisition system intended to be used in a hospital environment to record, collect, display and facilitate manual marking of aEEG recordings.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K123079, newest first; each opens FDA's recall record
May 7, 2025Z-1925-2025Electrode placement labels on the Digital Acquisition Box (DAB) were applied backwards. This may ultimately lead to misdiagnosis of seizure location in neonates monitored with this device.Open, Classified
Mar 5, 2020Z-1630-2020The OBM system functional test (impedance and noise) was not carried out on the affected items prior to release to the customer. The manufacturer cannot attest to the system functionality of the affected items.Terminated

Record

Type
Traditional
Received
Oct 1, 2012
Review panel
Neurology
Regulation
21 CFR 882.1400

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files