Olympic Brainz Monitor
Amplitude-Integrated Electroencephalograph. 510(k) K123079, from Excel-Tech Ltd. (Xltek) (Oakville, CA). Product code OMA, class 2.
As of Nov 17, 2026, K123079 is an FDA 510(k) held by Excel-Tech Ltd. (Xltek), substantially equivalent on May 8, 2013 after 219 days, with 2 recalls tied to it in FDA's recall database.
- Substantially equivalent: May 8, 2013
- Review time: 219 days
- Recalls: 2
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe Olympic Brainz Monitor (OBM) is a three channel electroencephalograph (EEG) acquisition system intended to be used in a hospital environment to record, collect, display and facilitate manual marking of aEEG recordings.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Olympic Brainz Monitor Model OBM70001K093949 · Natus Medical, Inc. · Jun 16, 2010
- Identevent, Version 1.0GK092039 · Optima Neuroscience, Inc. · Oct 16, 2009
- Persyst RevealK011397 · Persyst Development Corp. · Aug 3, 2001
Recalls
Recalls FDA ties to K123079, newest first; each opens FDA's recall record| May 7, 2025 | Z-1925-2025Electrode placement labels on the Digital Acquisition Box (DAB) were applied backwards. This may ultimately lead to misdiagnosis of seizure location in neonates monitored with this device. | Open, Classified |
| Mar 5, 2020 | Z-1630-2020The OBM system functional test (impedance and noise) was not carried out on the affected items prior to release to the customer. The manufacturer cannot attest to the system functionality of the affected items. | Terminated |
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Oct 1, 2012
- Review panel
- Neurology
- Regulation
- 21 CFR 882.1400
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files