MHI-TM2000 Linear Accelerator System Vero Imavis

Accelerator, Linear, Medical. 510(k) K122450, from Mitsubishi Heavy Industries, Ltd. (Hiroshima-Shi, JP). Product code IYE, class 2.

As of Nov 23, 2026, K122450 is an FDA 510(k) held by Mitsubishi Heavy Industries, Ltd., substantially equivalent on Oct 3, 2012 after 51 days, with 7 recalls tied to it in FDA's recall database.

  • Review time: 51 days
  • Recalls: 7
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

MHI-TM2000 Linear Accelerator System is intended for radiation therapy of lesions, tumors and conditions anywhere in the body where radiation treatment is indicated.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K122450, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Dec 8, 2017Z-1585-2018Due to a system controller software anomaly, the patient positioning deviation correction may not be applied and may result in the wrong part of the patient being irradiated.Terminated
Mar 7, 2016Z-1308-2016In rare cases the VERO/MHI-TM2000 Operator Console could set an incorrect Gantry angle for the first beam of an IMRT treatment if certain specific conditions are met. If Gantry angle for the first beam of IMRT is…Terminated
Dec 15, 2015Z-1244-2016Due to Operator Console software anomaly, a change in treatment completion status of the last patient of the day may be altered from "Completed (or Discontinued)" to "Untreated" under specific conditions.Terminated
Apr 17, 2015Z-1574-2015The operator console allows users to deliver therapeutic radiation to patients even though a specific communication error relevant to imaging conditions has occurred. In addition, the treatment record (the delivered…Terminated
Dec 2, 2014Z-0821-2015Software Anomaly. The VERO MHI-TM2000 Operator Console may provide Patient Positioning System (ExacTrac) with Image Angle Information Used for the 1st Port Image, for a subsequent port image fusion. As the result,…Terminated
Jul 5, 2014Z-2653-2014Software Anomaly: If a user changes the calendar setting from Workday to Holiday or vice versa, that would erroneously change the status of treatment/fractions, which are completed and to become Treated (completed)…Terminated
Jun 21, 2013Z-1931-2013Software anomaly: Mitsubishi Heavy Industries, Ltd. became aware of a potential safety issue associated with the VERO (TM) Linear Accelerator System. The VERO (TM) may save treatment delivery record into wrong location…Terminated

Other IYE clearances

The newest 510(k)s for the same product code; their median review was 173 days; see every IYE device

Record

Type
Traditional
Received
Aug 13, 2012
Review panel
Radiology
Regulation
21 CFR 892.5050

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files