Arca
Skip to content
Tools
Tools

Xhibit Central Station, Model 96102

Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms). 510(k) K122146, from Spacelabs Healthcare (Nederland, CO). Product code MHX, class 2.

As of Nov 19, 2026, K122146 is an FDA 510(k) held by Spacelabs Healthcare, substantially equivalent on Apr 22, 2013 after 277 days, with 5 recalls tied to it in FDA's recall database.

  • Review time: 277 days
  • Recalls: 5

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Spacelabs Healthcare Xhibit Central Station, Model 96102 intended use is to provide clinicians with central monitoring of adult, pediatric and neonatal patient data of patients connected to networked Spacelabs Healthcare patient monitors and telemetry transmitters.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K122146, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jun 4, 2026Z-2916-2026Due to off-the-shelf display port component that has audio/video failuresOpen
Aug 30, 2019Z-2261-2020The firm received reports of loss of audio alarm after a power failure or cable disconnection. The loss of audio for a telemetry patient could delay the recognition of an alarm condition.Terminated
Oct 18, 2016Z-0332-2017The firm received reports of telemetry SpO2 numerics dropping off the Xhibit Central display. Desaturation, high, and low limit alarms work normally.Terminated
Aug 25, 2016Z-2885-2016The firm received multiple reports of Xhibit Telemetry System going offline or locking up. In addition, short-duration asystole alarms may not display an audio and visual alarm as reported from a single Canadian…Terminated
Feb 26, 2016Z-1092-2016The firm has received one report of values for patient height and weight being switched when input at the Xhibit Central Station, Model 96102, causing a bedside monitor Body Surface Area (BSA) calculation to be in error.Terminated

Other MHX clearances

The newest 510(k)s for the same product code; their median review was 207 days; see every MHX device

Record

Type
Traditional
Received
Jul 19, 2012
Review panel
Cardiovascular
Regulation
21 CFR 870.1025

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc