Arca
Skip to content
Tools
Tools

Ivue With Normative Database

Ophthalmoscope, Ac-Powered. 510(k) K121739, from Optovue, Inc. (Fremont, CA). Product code HLI, class 2.

As of Nov 20, 2026, K121739 is an FDA 510(k) held by Optovue, Inc., substantially equivalent on Jan 18, 2013 after 219 days, with 3 recalls tied to it in FDA's recall database.

  • Review time: 219 days
  • Recalls: 3
  • Cited as a predicate by: 6 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The iVue is a non-contact, high resolution tomographic imaging device. It is intended for in vivo imaging, axial cross-sectional, and three-dimensional imaging and measurement of anterior and posterior ocular structures, including retina, retinal nerve fiber layer, ganglion cell complex (GCC), optic disc, cornea, and anterior chamber of the eye.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

6 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K121739, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jul 27, 2017Z-0008-2018FDA determined that a 510(k) is needed for the Vault Mapping software (also known as Lens Fitting).Terminated
Mar 16, 2016Z-1498-2016A software anomaly was detected that affects Optovue iVue and iVue 500 with Normative Database that may display images of the same eye rather than right and left if user scrolls on the map.Terminated
May 14, 2013Z-1441-2013Colors from the NDB comparison for Ganglion Cell Complex thickness in the singular OU report in iVue version 3.0 (US) and 3.1(international) are different from those in the individual iWellness ID and OS reports.Terminated

Record

Type
Traditional
Received
Jun 13, 2012
Review panel
Ophthalmic
Regulation
21 CFR 886.1570

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files