Laminar Flow Phaco Tips And Irrigation Sleeve (21G, Reusable)

Unit, Phacofragmentation. 510(k) K121721, from Abbott Medical Optics, Inc. (Santa Ana, CA). Product code HQC, class 2.

As of Nov 18, 2026, K121721 is an FDA 510(k) held by Abbott Medical Optics, Inc., substantially equivalent on Oct 15, 2012 after 125 days, with 1 recall tied to it in FDA's recall database.

  • Substantially equivalent: Oct 15, 2012
  • Review time: 125 days
  • Recalls: 1
  • Cited as a predicate by: 1 device

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The Laminar Flow Phacoemulsification Tips are used to break up (emulsify) the nucleus of a cataractous lens and remove the remaining nuclear fragments. The Irrigation Sleeve is a device intended to direct irrigation solution across the shaft of a phacoemulsification tip, allowing the solution to enter the eye during ocular surgery.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K121721, newest first; each opens FDA's recall record
StartedReasonStatus
Mar 14, 2024Z-1755-2024Due to a manufacturing variation of the irrigation sleeves which could result in a missing port hole, this non-conformity could lead to insufficient flow to cool the ultrasonic phaco tip, potentially leading to an…Open, Classified

Other HQC Clearances

The newest 510(k)s for the same product code; their median review was 267 days; see every HQC device

Record

Type
Traditional
Received
Jun 12, 2012
Review panel
Ophthalmic
Regulation
21 CFR 886.4670

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files