Accu-Chek Inform Ii Blood Glucose Monitoring System

System, Test, Blood Glucose, Over The Counter. 510(k) K121679, from Roche Diagnostics (Indianapolis, IN). Product code NBW, class 2.

As of Nov 19, 2026, K121679 is an FDA 510(k) held by Roche Diagnostics, substantially equivalent on Oct 11, 2012 after 126 days, with 3 recalls tied to it in FDA's recall database.

  • Review time: 126 days
  • Recalls: 3
  • Cited as a predicate by: 2 devices

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K121679, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jul 28, 2021Z-2354-2021Test strip vials may open while inside sealed cartons during shipment. An open vial might expose the test strips to humidity, which might damage the strips and result in inaccurate results, potentially leading to…Open
Feb 11, 2016Z-1172-2016Accu-Chek Inform II Base Unit might produce physical transmission errors in the form of data loss in the communication between the meter and the Data Management Systems (DMS). The issue can lead to the data loss or in…Terminated
Mar 6, 2013Z-1113-2013Roche investigation into complaints found Accu-Chek Inform II and CoaguChek XS Pro System barcode reading meters had the potential of intermittent erroneous decoding of patient identification if barcodes being read fail…Terminated

Other NBW clearances

The newest 510(k)s for the same product code; their median review was 213 days; see every NBW device

Record

Type
Traditional
Received
Jun 7, 2012
Review panel
Clinical Chemistry
Regulation
21 CFR 862.1345

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files