Centurion Vision System
Unit, Phacofragmentation. 510(k) K121555, from Alcon Laboratories, Inc. (Irvine, CA). Product code HQC, class 2.
As of Nov 18, 2026, K121555 is an FDA 510(k) held by Alcon Laboratories, Inc., substantially equivalent on Dec 21, 2012 after 210 days, with 4 recalls tied to it in FDA's recall database.
- Substantially equivalent: Dec 21, 2012
- Review time: 210 days
- Recalls: 4
- Cited as a predicate by: 2 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe CENTURION® Vision System is indicated for emulsification, separation, irrigation, and aspiration of cataracts, residual cortical material and lens epithelial cells, vitreous aspiration and cutting associated with anterior vitrectomy, bipolar coagulation, and intraocular lens injection.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Infiniti Vision System Evergreen Ii Autosert Iol Injector HandpieceK112425 · Alcon Manufacturing, Ltd. · Nov 21, 2011
- Constellation Vision SystemK101285 · Alcon Manufacturing, Ltd. · Nov 12, 2010
- Enhanced Ultra Vit ProbeK093305 · Alcon Research, Ltd. · Apr 2, 2010
- Alcon UltrachopperK091777 · Alcon Research, Ltd. · Oct 20, 2009
Cited as a Predicate
2 later 510(k) summaries name this device as a predicate- CENTURION Vision System (Active SentryTM)K161794 · Alcon Research, Ltd. · Apr 14, 2017
- Constellation Vision SystemK141065 · Alcon Research, Ltd. · Aug 22, 2014
Recalls
Recalls FDA ties to K121555, newest first; each opens FDA's recall record| Jun 28, 2023 | Z-2410-2023Reports of patient injury during use in procedures due to potential for plastic debris from over-tightening with included plastic wrench. | Open, Classified |
| Jun 28, 2023 | Z-2407-2023Reports of patient injury during use in procedures due to potential for plastic debris from over-tightening with included plastic wrench. | Open, Classified |
| May 24, 2016 | Z-2081-2016The grounding pin within the Power Entry Switch on certain systems was susceptible to detachment when the AC power cable is removed. | Terminated |
| Apr 28, 2016 | Z-1940-2016Alcon is conducting this Voluntary Medical Device Removal for specific lots of its CENTURION FMS packs due to the possible presence of a molding irregularity found within the aspiration luer which may result in a… | Terminated |
Other HQC Clearances
The newest 510(k)s for the same product code; their median review was 267 days; see every HQC device- MICOR 700 System (N/A); MICOR 700 Extractor (FG-50621)K262097 · Carl Zeiss Meditec Cataract Technologies · Oct 15, 2026
- Stellaris Elite vision enhancement system; Posterior and Combined Procedure Pack(s): 23 ga. Posterior Elite Pack…K261264 · Bausch and Lomb · May 15, 2026
- Stellaris Elite vision enhancement system (BL11145, BL14455, BL15455, SE14565, SE15565, SE14565E, SE15565E)K252052 · Bausch and Lomb, Incorporated · Mar 25, 2026
- System SophiK250501 · This AG · Nov 14, 2025
- MICOR 700 with Auto I/AK243395 · Carl Zeiss Meditec Cataract Technology, Inc. · Aug 8, 2025
- MICOR 700 System (N/A); MICOR 700 drive (FG-50631); MICOR 700 extractor (FG-50621); MICOR 700 vitrector (FG-51185)K242801 · Carl Zeiss Meditec Cataract Technology, Inc. · Jun 11, 2025
- Stellaris Elite vision enhancement system (BL11145, BL14455, BL15455)K240169 · Bausch and Lomb, Incorporated · Jul 26, 2024
- Faros Surgical SystemK233398 · Oertli Instrumente AG · Jun 27, 2024
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- May 25, 2012
- Review panel
- Ophthalmic
- Regulation
- 21 CFR 886.4670
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files